BD MAX¿ ENTERIC PARASITE PANEL
Report
- Report Number
- 3007420875-2022-00011
- Event Type
- Malfunction
- Date Received
- March 28, 2022
- Date of Event
- March 1, 2022
- Report Date
- May 10, 2022
- Manufacturer
- GENEOHM SCIENCES CANADA, INC. (BD DIAGNOSTICS)
- Product Code
- PCH
- UDI-DI
- 00382904429607
- PMA / PMN Number
- K143648
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INITIAL REPORTER PHONE#: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY THE COMPLAINT INVESTIGATION FOR ATYPICAL CURVES WITH THE KIT BD MAX¿ ENTERIC PARASITE PANEL (REF 442960) LOT 1103407 WAS PERFORMED BY THE REVIEW OF THE MANUFACTURING RECORDS, REVIEW OF CUSTOMER¿S DATA AND VERIFICATION OF COMPLAINTS HISTORY. REVIEW OF THE MANUFACTURING RECORDS OF BD MAX¿ ENTERIC PARASITE PANEL INDICATED THAT THE LOT WAS MANUFACTURED ACCORDING TO SPECIFICATIONS. CUSTOMER COMPLAINED ABOUT ATYPICAL CURVES THAT GAVE POSITIVE RESULTS WITH THE BD MAX¿ ENTERIC PARASITE PANEL (REF#442960) KIT LOT 1103407. CUSTOMER PROVIDED TWO RUNS FILE ((B)(4)) AND DATABASE FROM INSTRUMENT CT2261 FOR INVESTIGATION. ANALYSIS OF THE DATABASE REVEALED POSITIVE RESULTS FOR THE CRYPTO TARGET IN POSITIONS A1 AND A5 OF RUN (B)(4) AND POSITIONS A1, A2, A3, A4, A5 AND A7 OF RUN (B)(4), AS WELL AS POSITIVE RESULTS FOR THE GLAMB TARGET IN POSITIONS A1, A2 AND A3 OF RUN (B)(4), ALL ASSOCIATED TO UNUSUAL PCR CURVES THAT SUGGESTS THESE POSITIVE RESULTS ARE NOT TRUE AMPLIFICATIONS. REVIEW OF THE COMPLAINT DATABASE BY BD REVEALED THAT AN INSTRUMENT COMPLAINT WAS OPENED TO ASSESS THESE ATYPICAL CURVES. BASED ON THE CURRENT INVESTIGATION AND CURVES REVIEW, NO REAGENTS ISSUE IS SUSPECTED. THERE IS NO INDICATION OF A REAGENT ISSUE BASED ON THE ANALYSIS OF THE COMPLAINTS RECEIVED FOR ATYPICAL CURVES ON BD MAX¿ ENTERIC PARASITE PANEL KIT LOT 1103407. THE ROOT CAUSE IS STRONGLY SUSPECTED TO BE AN INSTRUMENT ISSUE. THE REAGENTS ARE NOT SUSPECTED OF BEING IN CAUSE. BD CANNOT CONFIRM THE COMPLAINT BASED ON THE INVESTIGATION THAT WAS PERFORMED. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED THAT WHILE USING THE BD MAX¿ ENTERIC PARASITE PANEL ON BD MAX¿ SYSTEM, A FALSE POSITIVE RESULT WAS OBTAINED. RESULTS WERE DEEMED UNRELIABLE WITHOUT CONFIRMATORY TESTING. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED, AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN TO ENGLISH: "VERY WEIRD CURVES CALLED POSITIVE."
IT WAS REPORTED THAT WHILE USING THE BD MAX¿ ENTERIC PARASITE PANEL ON BD MAX¿ SYSTEM, A FALSE POSITIVE RESULT WAS OBTAINED. RESULTS WERE DEEMED UNRELIABLE WITHOUT CONFIRMATORY TESTING. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED, AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN TO ENGLISH: "VERY WEIRD CURVES CALLED POSITIVE."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 298531 | BD MAX¿ ENTERIC PARASITE PANEL | GASTROINTESTINAL PATHOGEN PANEL MULTIPLEX NUCLEIC ACID-BASED ASSAY SYSTEM | PCH | GENEOHM SCIENCES CANADA, INC. (BD DIAGNOSTICS) | 442960 | 1103407 | 00382904429607 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |