FDA Adverse Event Injury Summary report: N

INCLUSIVE TAPERED IMPLANT 4.7 X 11.5 X 4.5MMP

MDR report key: 13929269 · Received March 28, 2022

Report

Report Number
3011649314-2022-00094
Event Type
Injury
Date Received
March 28, 2022
Date of Event
December 31, 2021
Report Date
June 26, 2024
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K121406
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HAS NOT BEEN RETURNED. IF/WHEN THERE IS MORE INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

CAPA 23-006. MANUFACTURER REFERENCE: (B)(4).

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED, THE INVESTIGATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS: THE DHR WAS REVIEWED FOR LOT# 6090266 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS: THE PACKAGED STOCK PRODUCT IS NOT APPLICABLE FOR REVIEW SINCE NO DEFECT WAS OBSERVED FROM THE RETURNED PRODUCT. INVESTIGATION METHODS/RESULTS: THE DEVICE WAS RETURNED BUT NOT IN ORIGINAL PACKAGE. THE DEVICE WAS VERIFIED TO BE AN INCLUSIVE TAPERED IMPLANT 4.7 MMD X 11.5 MML X 4.5 MMP (70-1070-IMP0012) USING RADIOGRAPHIC TEMPLATE (GD-437-091015). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. MINIMAL BONE DEBRIS WAS OBSERVED ON THE IMPLANT. THE COMPLAINT IS VERIFIED BASED ON THE RETURNED PART(S) BUT CANNOT CONFIRM THE FAILURE MODE. THERE WAS NO EVIDENCE FOUND THAT INDICATED THAT THE REPORTED ISSUE WAS CAUSED BY THE DEVICE ITSELF. ROOT CAUSE: "LACK OF PRIMARY STABILITY" IS A COMMON COMPLAINT IN REGARDS TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR LACK OF PRIMARY STABILITY IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. IFU 3023579 REV 3.0 (INCLUSIVE DENTAL IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." IN ADDITION. IFU 3023579 REV 3.0 (INCLUSIVE DENTAL IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN. ALSO, IFU 3023579 REV 3.0 (INCLUSIVE DENTAL IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN PRECAUTION SECTION: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL AND COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO. THIS COMPLAINT WILL BE KEPT ON RECORD FOR TRACK AND TRENDING PURPOSES.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE INCLUSIVE TAPERED IMPLANT FAILED TO INTEGRATE. HOWEVER, THE PIQ NOTES A LACK OF STABILITY, BUT BASED ON THE DATE OF IMPLANT TO EXPLANT IT WAS A FAILURE TO INTEGRATE. THE BONE GRADE IS NOTED AS GRADE IV. THE PATIENT HAS NO MEDICAL OR DENTAL HISTORY PRIOR TO IMPLANT. HOWEVER, IT IS NOTED THAT THE PATIENT IS A SMOKER. THE PATIENT PRESENTED ON (B)(6) 2021 FOR IMPLANT PLACEMENT ON TOOTH #3 AND 4. THERE THE PATIENT RETURNED ON (B)(6) 2021 BEFORE THE SECOND STAGE OF SURGERY WITHOUT COMPLAINTS. UPON EXAM THE PROVIDER NOTES A FAILURE TO INTEGRATE. IT WAS AT THAT TIME THE DEVICE WAS REMOVED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE INCLUSIVE TAPERED IMPLANT FAILED. THE PATIENT PRESENTED ON (B)(6) 2021 FOR IMPLANT PLACEMENT ON AN TOOTH #3. THE BONE GRADE IS NOTED AS GRADE IV. THE PATIENT RETURNED ON (B)(6) 2021 BEFORE THE SECOND STAGE OF SURGERY WITHOUT COMPLAINTS. UPON EXAM THE PROVIDER NOTES A LACK OF PRIMARY STABILITY. IT WAS AT THAT TIME THE DEVICE WAS REMOVED AND WAS NOT REPLACED. THERE WAS NO PERMANENT INJURY AND NO ADDITIONAL MEDICAL / SURGICAL PROCEDURE WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303949 INCLUSIVE TAPERED IMPLANT 4.7 X 11.5 X 4.5MMP INCLUSIVE TAPERED IMPLANT DZE PRISMATIK DENTALCRAFT, INC. 70-1070-IMP0012 6090266

Patients

Seq Age Sex Outcome Treatment
1 45 YR Male Required Intervention