QUICKSET ACE GRATER HEAD 56MM
Report
- Report Number
- 1818910-2022-05568
- Event Type
- Malfunction
- Date Received
- March 28, 2022
- Date of Event
- March 23, 2022
- Report Date
- March 28, 2022
- Manufacturer
- DEPUY ORTHOPAEDICS INC US
- Product Code
- HTO
- UDI-DI
- 10603295124030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY : THE DEVICE ASSOCIATED WITH THIS REPORT WAS RETURNED FOR ANALYSIS. EXAMINATION OF THE RETURNED INSTRUMENT CONFIRMED THE COMPLAINT. CUTTING SURFACE IS DULL TO THE TOUCH, NOT AS SHARP AS FIRST USE CONDITION. DEPUY SYNTHES CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION MAY BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT : THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED.
PRODUCT COMPLAINT #(B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H10 ADDITIONAL NARRATIVE: ADDED: D4 (LOT), D9, H4. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. CORRECTED: H3.
PRODUCT COMPLAINT (B)(4). INVESTIGATION SUMMARY: THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. DEPUY SYNTHES CONSIDERS, THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED. AND THE INVESTIGATION MAY BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT: A MANUFACTURING RECORD EVALUATION (MRE) WAS NOT POSSIBLE, BECAUSE THE REQUIRED LOT CODE WAS NOT PROVIDED.
IT WAS REPORTED THAT REAMERS BECOMING DULL..
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 797891 | QUICKSET ACE GRATER HEAD 56MM | HIP INSTRUMENTS : REAMERS | HTO | DEPUY ORTHOPAEDICS INC US | 2440-00-556 | SO2054263 | 10603295124030 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |