FDA Adverse Event Injury Summary report: N

TRIAGE METER PRO

MDR report key: 13924785 · Received March 26, 2022

Report

Report Number
3013982035-2022-00003
Event Type
Injury
Date Received
March 26, 2022
Date of Event
February 24, 2022
Report Date
March 25, 2022
Manufacturer
QUIDEL CARDIOVASCULAR INC.
Product Code
KHO
PMA / PMN Number
K973547
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION PENDING.

Additional Manufacturer Narrative · 0

THE TRIAGE METERPRO WAS INVESTIGATED IN-HOUSE WITH NO PROBLEM FOUND. THE CUSTOMER'S RETURNED POWER SUPPLY WAS SENT OUT TO THE VENDOR. THE VENDOR CONFIRMED THAT LIQUID RESIDUE LEAKED INTO THE POWER SUPPLY HOUSING CAUSING DAMAGE WHICH COULD LEAD TO ELECTRICAL FAILURES. THE ROOT CAUSE WAS DETERMINED TO BE ISSUE WITH FLUIDS GETTING INTO UNIT. NO CORRECTION ACTION IS REQUIRED.

Description of Event or Problem · 0

ON (B)(6) 2022: CUSTOMER REPORTED NURSE OBTAINED A SHOCK WHEN PLUGGING IN TRIAGE METER. TS REQUESTED ADDITIONAL INFORMATION TO UNDERSTAND "SHOCK" (IE. STATIC SHOCK OR ELECTRICAL). CUSTOMER ALSO STATED MACHINE BLEW UP BUT PHOTOS OF THE METER SHOW IT IS FULLY INTACT WITH NO SIGNS OF SMOKE OR HEAT DAMAGE. ON 03/3/22: TECHNICAL SUPPORT REQUESTED ADDITIONAL INFORMATION FROM THE ACCOUNT, THE NURSE RESPONDED WITH THE FOLLOWING: "I SUSTAINED AN ELECTRIC SHOCK TO MY LEFT HAND WHEN PLUGGING IN THE TRIAGE METER. MY FINGERS WERE PAINFUL FOR SOME TIME AFTERWARDS. I ALSO HAD A TINGLING PAIN FROM MY WRIST TO MY ELBOW. THIS RESOLVED COMPLETELY AFTER 3 DAYS. I DID NOT SEEK ANY MEDICAL ATTENTION. WITH HINDSIGHT I THINK I SHOULD HAVE VISITED A&E TO BE CHECKED. I COMPLETED AN INCIDENT REPORT AND WENT STRAIGHT OUT TO SEE MY PATIENTS."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
587942 TRIAGE METER PRO TRIAGE METER PRO KHO QUIDEL CARDIOVASCULAR INC. 55071

Patients

Seq Age Sex Outcome Treatment
1 Female Other