ETHILON NON ABS SUR SUT USP 4-0 70CM
Report
- Report Number
- 2210968-2022-02187
- Event Type
- Malfunction
- Date Received
- March 25, 2022
- Date of Event
- January 27, 2022
- Report Date
- May 19, 2022
- Manufacturer
- ETHICON INC.
- Product Code
- GAR
- PMA / PMN Number
- K946173
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
(B)(4). ADDITIONAL INFORMATION WAS REQUESTED, AND THE FOLLOWING WAS OBTAINED: HOW WAS THE PRODUCT PURCHASED? THE PRODUCT WAS PURCHASED DURING THE PROACTIVE GBP MARKET SURVEY USING BI APP. IS THERE AN INDICATION OF HOW THE PRODUCT WAS DISTRIBUTED? WE HAVE HANDED OVER THE LEAD TO GLOBAL SECURITY FOR INVESTIGATION ON THE IDENTIFIED STORES TO UNDERSTAND THE SUPPLY CHAIN AND DISTRIBUTION. WILL KEEP YOU UPDATED ONCE WE RECEIVE INFORMATION. IS THERE ANY INDICATION OF THE SOURCE? NO. BASED ON THE PACKAGING, IS THERE ANY INDICATION OF WHICH MARKET THE ORIGINAL GENUINE PRODUCT MAY HAVE COME FROM? THE ORIGINAL GENUINE PRODUCTS IS MANUFACTURED IN INDIA WAS THE DEVICE USED ON A PATIENT? IF SO, WAS THERE ANY PATIENT CONSEQUENCE? THE SAMPLE PRODUCTS ARE IN THE CUSTODY OF THE GBP VENDOR IT IS NOT BEEN USED ON THE PATIENT. INVESTIGATION SUMMARY: SAMPLE DETAILS: TWO PHOTOGRAPHS WERE PROVIDED OF PRODUCT CODE: NW3319 AND LOT NO: V8003. SUTURE MATERIAL: ETHILON (MONOFILAMENT POLYAMIDE BLACK). INVESTIGATION: RECEIVED SAMPLE WAS SUBJECTED TO INVESTIGATION IN ORDER TO IDENTIFY THE NOVELTY OF THE PRODUCT. FOLLOWING IS THE DETAIL OF THE EVALUATION PERFORMED DURING INVESTIGATION. VISUAL COMPARISON OF COMPLAINT SAMPLE WITH RETAIN SAMPLE: RECEIVED COMPLAINT SAMPLE PHOTOGRAPH WAS VISUALLY INSPECTED AND COMPARED WITH RETAIN SAMPLE OF PRODUCT CODE NW3319 AND LOT V8003. UPON VISUAL INSPECTION IT HAS BEEN OBSERVED THAT THE PRODUCT PACKING WAS NOT MATCHING WITH RETAINED SAMPLE. COMPLAINT SAMPLE SINGLE BARRIER FOLDER WAS COMPARED WITH RETAINED SAMPLE SINGLE BARRIER FOLDER AND DISCREPANCY WAS OBSERVED IN COMPLAINT SAMPLE WITH RESPECT TO ETHICON APPROVED ARTWORK, COLOR CODING, INK TYPE, PRODUCT ARTWORK WRITING STYLE FONT AND FOLDER CUTTING. IN ADDITION, 2D BARCODE SCANNING WAS PERFORMED OVER SINGLE BARRIER FOLDER AND NON-COMPLIANT RESULTS WERE OBSERVED. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT, AND NO NON-CONFORMANCES WERE IDENTIFIED. EVENT RELATED TO MW # 2210968-2022-02184, MW # 2210968-2022-02185.
PRODUCT COMPLAINT # (B)(4). H6 COMPONENT CODE: C22 - PHOTO ANALYSIS. DATE SENT TO FDA: 05/19/2022 THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. INVESTIGATION SUMMARY : INVESTIGATION: RECEIVED SAMPLE WAS SUBJECTED TO INVESTIGATION IN ORDER TO IDENTIFY THE NOVELTY OF THE PRODUCT. FOLLOWING IS THE DETAIL OF THE EVALUATION PERFORMED DURING INVESTIGATION. VISUAL COMPARISON OF COMPLAINT SAMPLE WITH RETAIN SAMPLE: RECEIVED COMPLAINT SAMPLE PHOTOGRAPH WAS VISUALLY INSPECTED AND COMPARED WITH RETAIN SAMPLE OF PRODUCT CODE NW3319 AND LOT B8003. UPON VISUAL INSPECTION IT HAS BEEN OBSERVED THAT THE PRODUCT PACKING WAS NOT MATCHING WITH RETAINED SAMPLE. COMPLAINT SAMPLE SINGLE BARRIER FOLDER WAS COMPARED WITH RETAINED SAMPLE SINGLE BARRIER FOLDER AND DISCREPANCY WAS OBSERVED IN COMPLAINT SAMPLE WITH RESPECT TO ETHICON APPROVED ARTWORK, COLOR CODING, INK TYPE, PRODUCT ARTWORK WRITING STYLE FONT AND FOLDER CUTTING. IN ADDITION, 2D BARCODE SCANNING WAS PERFORMED OVER SINGLE BARRIER FOLDER AND NON-COMPLIANT RESULTS WERE OBSERVED. BARCODE SCANNING DETAILS: BARCODE COMPLAINT SAMPLE SBF: (B)(4). RETAIN SAMPLE SBF:(B)(4). CONCLUSION: COMPLAINT SAMPLE AND RETAIN SAMPLE COMPARISON INCLUDING VISUAL INSPECTION REVEALED THAT COMPLAINT SAMPLE IS NOT A GENUINE PRODUCT MANUFACTURED AT ETHICON AURANGABAD INDIA SITE. THIS COMPLAINT IS CONCLUDED AS ¿CONFIRM COUNTERFEIT PRODUCT¿. CONSIDERING ABOVE INVESTIGATION ADEQUATE TO ADDRESS THE REPORTED COMPLAINT, THE COMPLAINT IS RECOMMENDED FOR CLOSURE APPROVAL.
IT WAS REPORTED THAT GLOBAL BRAND PROTECTION HAD CONDUCTED A COVERT MARKET SURVEY FOR SUTURES IN CITY OF (B)(6) OCCURRED IN THE MONTH OF (B)(6) 2022. DURING THE MARKET SURVEY, IT WAS IDENTIFIED THAT (B)(6) IS INVOLVED IN THE SALE OF COUNTERFEIT SUTURE PRODUCTS. A SAMPLE WAS PROCURED FROM (B)(6) AND SURGICAL. ADDITIONAL INFORMATION WAS REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1054841 | ETHILON NON ABS SUR SUT USP 4-0 70CM | SUTURE, NONABSORBABLE, SYNTHETIC | GAR | ETHICON INC. | B8003 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |