FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY

MDR report key: 13891396 · Received March 24, 2022

Report

Report Number
3013756811-2022-29257
Event Type
Malfunction
Date Received
March 24, 2022
Date of Event
March 6, 2022
Report Date
March 6, 2022
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
UDI-DI
00850006613779
PMA / PMN Number
K201214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HAS BEEN RECEIVED FOR EVALUATION; HOWEVER, DEVICE EVALUATION IS NOT YET COMPLETE. A SUPPLEMENTAL REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT CARTRIDGE CHANGE ERRORS OCCURRED WITH MULTIPLE CARTRIDGES. ADDITIONALLY, AN O-RING WAS ON THE PNEUMATIC TAP. THE CUSTOMER WAS ABLE TO REMOVE THE O-RING FROM THE PUMP. SUBSEQUENTLY, PUMP DECLARED A MINIMUM FILL NOTIFICATION OCCURRED AFTER FILLING THE CARTRIDGE WITH 300 UNITS OF INSULIN. CUSTOMER¿S BLOOD GLUCOSE RANGED FROM 160-177 MG/DL. CUSTOMER REVERTED TO MANUAL INJECTION FOR INSULIN THERAPY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
367532 T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1002717 00850006613779

Patients

Seq Age Sex Outcome Treatment
1 74 YR Female