FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 13890450 · Received March 24, 2022

Report

Report Number
1221359-2022-01670
Event Type
Malfunction
Date Received
March 24, 2022
Date of Event
March 4, 2022
Report Date
April 13, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC
Product Code
QJR
UDI-DI
10811877011269
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. REFERENCE MFR. REPORTS: 1221359-2022-01647 THROUGH 1221359-2022-01672 .

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1051611 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT 1051611, TEST BASE PART NUMBER 190-430 / LOT 1051611. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1051611 SHOWED THAT THE COMPLAINT RATE IS (B)(4). A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1051611 SHOWED THAT THE COMPLAINT RATE IS 0.048% (26/53760). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE AS THE LOGFILES WERE NOT PROVIDED. HOWEVER, A POSSIBLE ROOT CAUSE IS CROSS CONTAMINATION.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TWENTY-SIX (26) CONFLICTING RESULTS WITH THE IDNOW COVID-19 PERFORMED ON VARIOUS DATES INVOLVING LOT NUMBER (1051611) TOTAL QUANTITY (1). THIS REPORT ADDRESSES TEST TWENTY FOUR (24) PERFORMED ON (B)(6) 2022 OF TWENTY-SIX (26) TESTS. CONFIRMATION TESTING WAS NOT PERFORMED. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED. NO INFORMATION WAS PROVIDED IF PATIENT WAS ASYMPTOMATIC OR SYMPTOMATIC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
797247 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC 1051611 10811877011269

Patients

Seq Age Sex Outcome Treatment
1 Unknown