ID NOW COVID-19 ASSAY
Report
- Report Number
- 1221359-2022-01661
- Event Type
- Malfunction
- Date Received
- March 24, 2022
- Date of Event
- March 3, 2022
- Report Date
- April 11, 2022
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC
- Product Code
- QJR
- UDI-DI
- 10811877011269
- PMA / PMN Number
- EUA2000074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. REFERENCE MFR. REPORTS: 1221359-2022-01647 THROUGH 1221359-2022-01672.
TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1051611 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT 1051611, TEST BASE PART NUMBER 190-430 / LOT 1051611. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE AS THE LOGFILES WERE NOT PROVIDED. HOWEVER, A POSSIBLE ROOT CAUSE IS CROSS CONTAMINATION.
THE CUSTOMER REPORTED TWENTY-SIX (26) CONFLICTING RESULTS WITH THE IDNOW COVID-19 PERFORMED ON VARIOUS DATES INVOLVING LOT NUMBER (1051611) TOTAL QUANTITY (1). THIS REPORT ADDRESSES TEST FIFTEEN (15) PERFORMED ON (B)(6) 2022 OF TWENTY-SIX (26) TESTS. CONFIRMATION TESTING WAS NOT PERFORMED. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED. NO INFORMATION WAS PROVIDED IF PATIENT WAS ASYMPTOMATIC OR SYMPTOMATIC.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 864144 | ID NOW COVID-19 ASSAY | MOLECULAR IVD FOR ID NOW COVID-19 | QJR | ABBOTT DIAGNOSTICS SCARBOROUGH, INC | 1051611 | 10811877011269 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |