FDA Adverse Event Injury Summary report: N

GAP SPACER

MDR report key: 13870475 · Received March 23, 2022

Report

Report Number
1020279-2022-01320
Event Type
Injury
Date Received
March 23, 2022
Date of Event
March 2, 2022
Report Date
April 8, 2022
Manufacturer
SMITH & NEPHEW, INC.
Product Code
HWT
UDI-DI
00885556578445
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INTERNAL COMPLAINT REFERENCE: (B)(4).

Additional Manufacturer Narrative · 0

THE ASSOCIATED DEVICE WAS RETURNED AND EVALUATED. A VISUAL INSPECTION OF THE RETURNED DEVICE CONFIRMED THE STATED FAILURE MODE. THE GAP SPACER IS BROKEN INTO 2 SEPARATE PIECES, RENDERING THE DEVICE INOPERATIVE. BOTH PIECES WERE RETURNED. A REVIEW OF COMPLAINT HISTORY BASED ON THE HISTORICAL DATA REVEALED SIMILAR EVENTS FOR THE LISTED BATCH, THIS FAILURE MODE WILL BE MONITORED FOR FUTURE COMPLAINTS FOR ANY NECESSARY CORRECTIVE ACTIONS. A REVIEW OF THE MANUFACTURING RECORDS DID NOT REVEAL A MANUFACTURING ABNORMALITY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. AT THIS TIME, WE DO NOT HAVE REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. A REVIEW OF THE RISK MANAGEMENT FILE REVEALED THIS FAILURE MODE WAS PREVIOUSLY IDENTIFIED. THE ANTICIPATED RISK LEVEL IS STILL ADEQUATE. ASSESSMENT OF HISTORICAL ESCALATED CASES CONCLUDED THAT THERE ARE NO PRIOR ACTIONS RELATED TO THIS DEVICE AND FAILURE MODE. AS THE DEVICE BROKE AND IT CANNOT LONGER FIT ITS PURPOSE, THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD BE CORROBORATED. THIS DEVICE IS A REUSABLE INSTRUMENT THAT CAN BE EXPOSED TO NUMEROUS SURGERIES. DAMAGE FROM PROLONGED USE, MISUSE OR ROUGH HANDLING ARE PROBABLE CAUSES OF THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THIS COMPLAINT; HOWEVER, WE WILL CONTINUE TO MONITOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY. WE CONSIDER THIS INVESTIGATION CLOSED.

Description of Event or Problem · 0

IT WAS REPORTED THAT, DURING A TKA SURGERY, THE GAP SPACER BROKE INSIDE PATIENT. ALL PIECES WERE RETRIEVED. SURGERY WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED BEYOND THE REPORTED PROBLEM. NO FURTHER COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1070091 GAP SPACER TEMPLATE HWT SMITH & NEPHEW, INC. 71935180 19GDN3321 00885556578445

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention