FDA Adverse Event Malfunction Summary report: N

JOURNEY FEM IMPL IMPACTOR

MDR report key: 13870341 · Received March 23, 2022

Report

Report Number
1020279-2022-01317
Event Type
Malfunction
Date Received
March 23, 2022
Date of Event
March 2, 2022
Report Date
April 20, 2022
Manufacturer
SMITH & NEPHEW, INC.
Product Code
HRY
UDI-DI
03596010556905
PMA / PMN Number
K121393
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4).

Additional Manufacturer Narrative · 0

H3, H6: THE DEVICE WAS NOT RETURNED FOR EVALUATION AND THE REPORTED EVENT COULD NOT BE CONFIRMED. A REVIEW OF THE MANUFACTURING RECORDS DID NOT REVEAL A MANUFACTURING ABNORMALITY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. A REVIEW OF COMPLAINT HISTORY REVEALED SIMILAR EVENTS FOR THE LISTED DEVICE OVER THE PREVIOUS 12 MONTHS, BUT NO SIMILAR EVENTS FOR THE BATCH BASED ON THE HISTORICAL DATA, THIS FAILURE MODE WILL BE MONITORED FOR FUTURE COMPLAINTS FOR ANY NECESSARY CORRECTIVE ACTIONS. A REVIEW OF THE RISK MANAGEMENT FILE REVEALED THIS FAILURE MODE WAS PREVIOUSLY IDENTIFIED. THE ANTICIPATED RISK LEVEL IS STILL ADEQUATE. A HISTORICAL REVIEW CONCLUDED THAT THERE ARE NO PRIOR ACTIONS RELATED TO THIS PRODUCT AND EVENT. AT THIS TIME, WE HAVE NO REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. THIS DEVICE IS A REUSABLE INSTRUMENT THAT CAN BE EXPOSED TO NUMEROUS SURGERIES. DAMAGE FROM PROLONGED USE, MISUSE OR ROUGH HANDLING ARE LIKELY PROBABLE CAUSES OF THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD NOT BE CORROBORATED. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. WITHOUT THE RETURN OF THE ACTUAL PRODUCT INVOLVED, OUR INVESTIGATION COULD NOT PROCEED. SHOULD THE DEVICE OR ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THIS COMPLAINT; HOWEVER, WE WILL CONTINUE TO MONITOR FOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY. WE CONSIDER THIS INVESTIGATION CLOSED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE GREY PLASTIC OF A JOURNEY FEM IMPL IMPACTOR IS CHIPPING OFF. AS THIS WAS NOTICED DURING SET UP FOR A TKA, THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT IMPACTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
714690 JOURNEY FEM IMPL IMPACTOR PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER HRY SMITH & NEPHEW, INC. 74012831 14FM04262 03596010556905

Patients

Seq Age Sex Outcome Treatment
1 Unknown