FDA Adverse Event Malfunction Summary report: N

PIC IX HARDWARE

MDR report key: 13862021 · Received March 23, 2022

Report

Report Number
1218950-2022-00239
Event Type
Malfunction
Date Received
March 23, 2022
Date of Event
February 25, 2022
Manufacturer
PHILIPS NORTH AMERICA LLC
Product Code
MHX
PMA / PMN Number
K102495
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETE.

Additional Manufacturer Narrative · 0

A PHILIPS REMOTE CLINICAL SPECIALIST (RCS) SPOKE WITH THE CUSTOMER. THE RCS REVIEWED THE ALARM REVIEW . NO PATIENT STRIP WAS PROVIDED. THE RCS NOTED AT 09:21:24 THE SPO2 PARAMETER WAS TURNED OFF AND IT WAS TURNED BACK ON AT 09:21:38 ON (B)(6) 2022. THE NEXT ENTRY IS DESATURATION OF 57% AT 09:56 WHICH ENDED AT 09:57. THERE WAS NO SPO2 ALARM ENTRY FOUND AT APPROXIMATELY 09:30 HOURS. ON (B)(6) 2022 10:33:00, (B)(4), DESAT 76 < 80 GENERATED AT 10:32:47, PIC IX: WRXPCU. ON (B)(6) 2022 10:32:43,(B)(4), DESAT 76 < 80 GENERATED AT 10:32:47, PIC IX: PCUCLNT. ON (B)(6) 2022 10:29:55, (B)(4), SILENCE, WRXPCU. ON (B)(6) 2022 10:27:14, (B)(4), SILENCE, WRXPCU. ON (B)(6) 2022 10:12:44, (B)(4), SILENCE, WRXPCU. ON (B)(6) 2022 09:57:34, (B)(4) , PULSE -PULSE FROM OFF TO ON, MPCU6. ON (B)(6) 2022 09:57:34, (B)(4), SPO2 FROM OFF TO ON, MPCU6. ON (B)(6) 2022 09:57:21, (B)(4), PULSE -PULSE FROM OFF TO ON, MPCU6. ON (B)(6) 2022 09:57:21, (B)(4), SILENCE, MPCU6. ON (B)(6) 2022 09:57:21, (B)(4), SPO2 - FROM ON TO OFF, MPCU6. ON (B)(6) 2022 09:57:20, (B)(4), DESAT 57 < 80 ENDED, PIC IX: WRXPCU. ON (B)(6) 2022 09:57:04, (B)(4), DESAT 57 < 80 ENDED, PIC IX: PCUCLNT. ON (B)(6) 2022 09:57:20, (B)(4), SILENCE MPCU6. ON (B)(6) 2022 09:57:03, (B)(4), DESAT 57 < 80 GENERATED AT 09:56:52, PIC IX: WRXPCU. ON (B)(6) 2022 09:56:46, (B)(4), DESAT 57 < 80 GENERATED AT 09:56:52, PIC IX: PCUCLNT. ON (B)(6) 2022 09:21:38, (B)(4), PULSE -PULSE FROM OFF TO ON, MPCU6. ON (B)(6) 2022 09:21:38,(B)(4), SPO2- FROM OFF TO ON, MPCU6. ON (B)(6) 2022 09:21:25, (B)(4), SPO2 - ALARMS FROM OFF TO ON, MPCU6. ON (B)(6) 2022 09:21:24, (B)(4), PULSE -PULSE FROM ON TO OFF, MPCU6. ON (B)(6) 2022 09:21:24, (B)(4), SPO2 - FROM ON TO OFF, MPCU6. ON (B)(6) 2022 09:20:51, (B)(4), SILENCE, MPCU6. ON (B)(6) 2022 09:07:07, (B)(4), SILENCE, MPCU6. ON (B)(6) 2022 08:44:53, (B)(4), SILENCE, MPCU6. ON (B)(6) 2022 08:24:26, (B)(4), SILENCE, MPCU6. ON (B)(6) 2022 08:09:00, (B)(4), SILENCE, WRXPCU. ON (B)(6) 2022 08:08:58, (B)(4), SILENCE, WRXPCU. ON (B)(6) 2022 07:22:39, (B)(4), SILENCE, WRXPCU. A GOOD FAITH EFFORT WAS MADE TO OBTAIN ADDITIONAL INFORMATION ASSOCIATED WITH THIS COMPLAINT EVALUATION, BUT ATTEMPTS HAVE BEEN UNSUCCESSFUL. IF ADDITIONAL INFORMATION IS LATER OBTAINED, THE COMPLAINT WILL BE REOPENED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT ON (B)(6) 2022 AT APPROX. 0930 HOURS, THE ROOM #PCU 6 PATIENT HAD A DESATURATION EVENT OF 65% AND IT DID NOT ALARM. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT ON (B)(6) 2022 AT APPROX. 0930 HOURS, THE ROOM#: (B)(4) PATIENT HAD A DESATURATION EVENT OF 65% AND IT DID NOT ALARM THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
562192 PIC IX HARDWARE PIC IX HARDWARE MHX PHILIPS NORTH AMERICA LLC 866424

Patients

Seq Age Sex Outcome Treatment
1 Unknown