FDA Adverse Event Malfunction Summary report: N

L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM

MDR report key: 13860404 · Received March 23, 2022

Report

Report Number
0001625425-2022-00967
Event Type
Malfunction
Date Received
March 23, 2022
Date of Event
February 25, 2022
Report Date
April 25, 2022
Manufacturer
ARGON MEDICAL DEVICES
Product Code
FOZ
UDI-DI
00886333209934
PMA / PMN Number
K091670
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE SAMPLE IS INDICATED AS AVAILABLE FOR RETURN. AS OF THE DATE OF THIS REPORT, THE SAMPLE HAS NOT BEEN RETURNED. A FOLLOW-UP REPORT WILL BE PROVIDED ONCE THE DEVICE HAS BEEN RECEIVED AND REVIEWED.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DHR AND INSPECTION RECORDS WAS CONDUCTED, AND NO SIMILAR CONCERNS WERE FOUND. ONE SAMPLE WAS RETURNED FOR REVIEW. VISUAL INSPECTION OF THE RETURNED SAMPLE CONFIRMED A CRACK IN THE HUB WHICH COULD RESULT IN LEAKAGE. SEVERAL COMPLAINTS HAVE PREVIOUSLY BEEN RECEIVED REGARDING A CRACKED HUB RESULTING IN LEAKAGE, AND CAPA 2021-039 HAS BEEN INITIATED TO ADDRESS THIS ISSUE AND IS CURRENTLY UNDER INVESTIGATION. THE CAPA WILL IDENTIFY THE SPECIFIC CAUSES AND CORRECTIVE ACTIONS TO BE TAKEN. CAPA 2021-039 HAS BEEN INITIATED TO ADDRESS THIS ISSUE AND IS CURRENTLY UNDER INVESTIGATION. THE CAPA WILL IDENTIFY THE SPECIFIC CAUSES AND CORRECTIVE ACTIONS TO BE TAKEN.

Description of Event or Problem · 0

PICC LINE PLACED AS REPLACEMENT PICC ON PER FORM (B)(6) PENN HOSPITAL #2 HUB CRACKED AND LEAKED AT 1755PM ON (B)(6) 2022. PER CUSTOMER APPROXIMATELY 23 HOURS AFTER INSERTION. PICC D/C¿ED. PIV PLACED BECAUSE PATIENT ON TPN, LIPIDS, AND ANTIBIOTICS THIS IS THE SAME PATIENT AS AGH WEST PENN HOSPITAL PER FORM #2. PATIENT HAD TWO PICC INSERTIONS WITHIN APPROXIMATELY 23 HOURS.

Description of Event or Problem · 0

PICC LINE PLACED AS REPLACEMENT PICC ON PER FORM AGH WEST PENN HOSPITAL #2 HUB CRACKED AND LEAKED AT 1755PM ON 3/13/2022. PER CUSTOMER APPROXIMATELY 23 HOURS AFTER INSERTION. PICC D/C¿ED. PIV PLACED BECAUSE PATIENT ON TPN, LIPIDS, AND ANTIBIOTICS THIS IS THE SAME PATIENT AS AGH WEST PENN HOSPITAL PER FORM #2. PATIENT HAD TWO PICC INSERTIONS WITHIN APPROXIMATELY 23 HOURS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2365870 L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM L-CATH PICC FOZ ARGON MEDICAL DEVICES 384539 11403016 00886333209934

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other