L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM
Report
- Report Number
- 0001625425-2022-00967
- Event Type
- Malfunction
- Date Received
- March 23, 2022
- Date of Event
- February 25, 2022
- Report Date
- April 25, 2022
- Manufacturer
- ARGON MEDICAL DEVICES
- Product Code
- FOZ
- UDI-DI
- 00886333209934
- PMA / PMN Number
- K091670
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
THE SAMPLE IS INDICATED AS AVAILABLE FOR RETURN. AS OF THE DATE OF THIS REPORT, THE SAMPLE HAS NOT BEEN RETURNED. A FOLLOW-UP REPORT WILL BE PROVIDED ONCE THE DEVICE HAS BEEN RECEIVED AND REVIEWED.
A REVIEW OF THE DHR AND INSPECTION RECORDS WAS CONDUCTED, AND NO SIMILAR CONCERNS WERE FOUND. ONE SAMPLE WAS RETURNED FOR REVIEW. VISUAL INSPECTION OF THE RETURNED SAMPLE CONFIRMED A CRACK IN THE HUB WHICH COULD RESULT IN LEAKAGE. SEVERAL COMPLAINTS HAVE PREVIOUSLY BEEN RECEIVED REGARDING A CRACKED HUB RESULTING IN LEAKAGE, AND CAPA 2021-039 HAS BEEN INITIATED TO ADDRESS THIS ISSUE AND IS CURRENTLY UNDER INVESTIGATION. THE CAPA WILL IDENTIFY THE SPECIFIC CAUSES AND CORRECTIVE ACTIONS TO BE TAKEN. CAPA 2021-039 HAS BEEN INITIATED TO ADDRESS THIS ISSUE AND IS CURRENTLY UNDER INVESTIGATION. THE CAPA WILL IDENTIFY THE SPECIFIC CAUSES AND CORRECTIVE ACTIONS TO BE TAKEN.
PICC LINE PLACED AS REPLACEMENT PICC ON PER FORM (B)(6) PENN HOSPITAL #2 HUB CRACKED AND LEAKED AT 1755PM ON (B)(6) 2022. PER CUSTOMER APPROXIMATELY 23 HOURS AFTER INSERTION. PICC D/C¿ED. PIV PLACED BECAUSE PATIENT ON TPN, LIPIDS, AND ANTIBIOTICS THIS IS THE SAME PATIENT AS AGH WEST PENN HOSPITAL PER FORM #2. PATIENT HAD TWO PICC INSERTIONS WITHIN APPROXIMATELY 23 HOURS.
PICC LINE PLACED AS REPLACEMENT PICC ON PER FORM AGH WEST PENN HOSPITAL #2 HUB CRACKED AND LEAKED AT 1755PM ON 3/13/2022. PER CUSTOMER APPROXIMATELY 23 HOURS AFTER INSERTION. PICC D/C¿ED. PIV PLACED BECAUSE PATIENT ON TPN, LIPIDS, AND ANTIBIOTICS THIS IS THE SAME PATIENT AS AGH WEST PENN HOSPITAL PER FORM #2. PATIENT HAD TWO PICC INSERTIONS WITHIN APPROXIMATELY 23 HOURS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2365870 | L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM | L-CATH PICC | FOZ | ARGON MEDICAL DEVICES | 384539 | 11403016 | 00886333209934 |