FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 13847663 · Received March 22, 2022

Report

Report Number
1000113657-2022-00153
Event Type
Malfunction
Date Received
March 22, 2022
Date of Event
February 22, 2022
Report Date
May 3, 2022
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. MANUFACTURER CONTACTED CUSTOMER IN SEVERAL FOLLOW-UP CALLS TO ENSURE THE INITIAL CONCERN IS RESOLVED - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME.

Additional Manufacturer Narrative · 0

SECTIONS WITH ADDITIONAL INFORMATION AS OF 03-MAY-2022: H6: UPDATED FDA'S TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS. H10: METER AND STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIPS TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-058: USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 183, 167, 251 AND 208MG/DL. THE CUSTOMER¿S EXPECTED PM FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 120-130MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE LIVING ROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 04/30/2023 AND OPEN VIAL DATE IS ONE WEEK. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 183 MG/DL; DATE: (B)(6) 2022; TIME: 3:40PM; FASTING. RESULT 2: 167 MG/DL; DATE: (B)(6) 2022; TIME: 10:20PM; FASTING. RESULT 3: 229 MG/DL; DATE: (B)(6) 2022; TIME: 10:30PM; FASTING. RESULT 4: 251 MG/DL; DATE: (B)(6) 2022; TIME: 4:30PM; FASTING. RESULT 5: 208 MG/DL; DATE: (B)(6) 2022; TIME: 4:30PM; FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
264586 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, SUNMARK TMX 50CTMG/DL MM ZY4532S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1 Unknown