TRUE METRIX
Report
- Report Number
- 1000113657-2022-00151
- Event Type
- Malfunction
- Date Received
- March 21, 2022
- Date of Event
- February 22, 2022
- Report Date
- April 27, 2022
- Manufacturer
- TRIVIDIA HEALTH INC
- Product Code
- NBW
- UDI-DI
- 00021292007836
- PMA / PMN Number
- K140100
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
INTERNAL REPORT REFERENCE NUMBER: 106098-1. METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON (B)(6) 2022 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN -ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED REPLACEMENT PRODUCTS RESOLVED INITIAL CONCERN.
SECTIONS WITH ADDITIONAL INFORMATION AS OF 27-APR-2022: D9: DEVICE AVAILABLE FOR EVALUATION. H3: DEVICE EVALUATED BY MANUFACTURER. H6: UPDATED FDA¿S TYPE, FINDINGS AND CONCLUSIONS CODES. H10: METER AND TEST STRIPS WERE RETURNED FOR EVALUATION. PRODUCT TESTING WAS PERFORMED AND NO DEFECT FOUND. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIP LOT TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-018: USER HAS HIGH GLUCOSE VALUE.
CONSUMER REPORTED COMPLAINT FOR HI, HIGH AND ERRATIC BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF HI, 114, 127, 115, 480MG/DL. THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 60-90MG/DL AND EXPECTED ONE HOUR POST MEAL EXPECTED BLOOD GLUCOSE TEST RESULT RANGE IS 100-130MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED TEST RESULTS OF 139MG/DL AND 140MG/DL USING TRUE METRIX METER.; THE CUSTOMER WAS SATISFIED WITH THE RESULTS OBTAINED. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE DINING ROOM/LIVING ROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS (B)(6) 2023 AND OPEN VIAL DATE IS (B)(6) 2022. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY:RESULT 1: 114MG/DL DATE: (B)(6) 2022 TIME: 8:06 PM 1 HR. AFTER MEAL. RESULT 2: 127MG/DL DATE: (B)(6) 2022 TIME: 8:04 PM 1 HR. AFTER MEAL.RESULT 3: 115MG/DL DATE: (B)(6) 2022 TIME: 1:32 PM 1 HR. AFTER MEAL.RESULT 4: 480MG/DL DATE: (B)(6) 2022 TIME: 7:31 AM FASTING.RESULT 5: HI MG/DL DATE: (B)(6) 2022 TIME: 7:30 AM FASTING.