BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE
Report
- Report Number
- 3003152976-2022-00109
- Event Type
- Malfunction
- Date Received
- March 21, 2022
- Date of Event
- February 8, 2022
- Report Date
- April 28, 2022
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVAL YES, D9: RETURNED TO MANUFACTURER ON: 21-MAR-2022. H6: INVESTIGATION SUMMARY: TWO SAMPLES WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, A TRANSPARENT RESIDUE WAS OBSERVED ON THE STOPPER OF THE SYRINGES, A HIGHER QUANTITY NOTED IN ONE OF THE SAMPLES. THROUGH OUR EVALUATION, THE RESIDUE WAS IDENTIFIED TO BE SILICONE. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2110136, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. LUBRICANT IS EMPLOYED DURING THE SYRINGE ASSEMBLY PROCESS TO LUBRICATE THE CYLINDERS IN THE SILICONE STATION. THE SILICONE EMPLOYED IN THIS PRODUCT IS A MEDICAL GRADE SILICONE AUTHORIZED FOR PRODUCT USE. SILICONE CONTENT TESTS ARE PERFORMED DURING THE MANUFACTURING PROCESS OF EACH LOT NUMBER. RESULTS WERE REVIEWED FOR LOT 2110136 AND FOUND TO BE WITHIN SPECIFICATION. THE RETURNED SAMPLES UNDERWENT THE SAME TESTING, VERIFYING ONE SAMPLE WAS ABOVE THE SPECIFIED LIMIT, THE SECOND SYRINGE WAS WITHIN REQUIRED LIMITS.
IT WAS REPORTED THAT GREASE-LIKE FOREIGN MATTER WAS FOUND ON 2 BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGES' PLUNGER TIPS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6): "PROBLEM OBSERVED: INSIDE THE SYRINGE, ON THE TIP OF THE PLUNGER, DEPOSIT OF A LAYER OF GREASE IMMEDIATE ACTION: DEVICE CHANGE".
IT WAS REPORTED THAT GREASE-LIKE FOREIGN MATTER WAS FOUND ON 2 BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGES' PLUNGER TIPS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH: "PROBLEM OBSERVED: INSIDE THE SYRINGE, ON THE TIP OF THE PLUNGER, DEPOSIT OF A LAYER OF GREASE IMMEDIATE ACTION: DEVICE CHANGE."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 651155 | BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 2110136 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |