FDA Adverse Event Malfunction Summary report: N

ID NOW COVID 19 ASSAY

MDR report key: 13832063 · Received March 20, 2022

Report

Report Number
1221359-2022-01709
Event Type
Malfunction
Date Received
March 20, 2022
Date of Event
February 26, 2022
Report Date
March 19, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
10811877011320
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(6), INC. ON RETAINED KIT LOT 1060760 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 191-000 / LOT: 1060760, TEST BASE PART NUMBER 190-430/ LOT: 1060760. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE AND FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1060760 SHOWED THAT THE COMPLAINT RATE IS 0.020%. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION. REFERENCE MFR REPORT NUMBERS: 1221359-2022-01702 THROUGH 1221359-2022-01711.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TEN (10) CONFLICTING RESULTS WITH THE ID NOW COVID-19 ASSAY FOR TEN (10) PATIENTS PERFORMED ON (B)(6) 2022. THIS MFR. REPORT ADDRESSES PATIENT EIGHT (8) AND IS EIGHT (8) OF TEN (10) . THE CUSTOMER REPORTED A CONFLICTING RESULT WITH THE ID NOW COVID-19 ASSAY TESTED ON A NASAL KITTED SWAB. CUSTOMER MENTIONED SOME PATIENTS WERE CLOSE CONTACT AND HAD SYMPTOMS. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
651611 ID NOW COVID 19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1060760 10811877011320

Patients

Seq Age Sex Outcome Treatment
1 Unknown