ID NOW COVID 19 ASSAY
Report
- Report Number
- 1221359-2022-01711
- Event Type
- Malfunction
- Date Received
- March 20, 2022
- Date of Event
- February 26, 2022
- Report Date
- March 19, 2022
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QJR
- UDI-DI
- 10811877011320
- PMA / PMN Number
- EUA2000074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RE
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT 1060760 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 191-000 / LOT: 1060760, TEST BASE PART NUMBER 190-430/ LOT: 1060760. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE AND FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1060760 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION. REFERENCE MFR REPORT NUMBERS: 1221359-2022-01702 THROUGH 1221359-2022-01710.
THE CUSTOMER REPORTED TEN (10) CONFLICTING RESULTS WITH THE ID NOW COVID-19 ASSAY FOR TEN (10) PATIENTS PERFORMED ON (B)(6) 2022. THIS MFR. REPORT ADDRESSES PATIENT TEN (10) AND IS TEN (10) OF TEN (10). THE CUSTOMER REPORTED A CONFLICTING RESULT WITH THE ID NOW COVID-19 ASSAY TESTED ON A NASAL KITTED SWAB. CUSTOMER MENTIONED SOME PATIENTS WERE CLOSE CONTACT AND HAD SYMPTOMS. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 651610 | ID NOW COVID 19 ASSAY | MOLECULAR IVD FOR ID NOW COVID-19 | QJR | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 1060760 | 10811877011320 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |