FDA Adverse Event Injury Summary report: N

BALL HEADS: MECTACER 01.29.202 BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 28 SIZE M

MDR report key: 13826438 · Received March 19, 2022

Report

Report Number
3005180920-2022-00181
Event Type
Injury
Date Received
March 19, 2022
Date of Event
February 17, 2022
Report Date
March 18, 2022
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030809132
PMA / PMN Number
K112115
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 18 FEBRUARY 2022. LOT 2012905: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 09-APR-2021. EXPIRATION DATE: 2026-03-25. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. OTHER DEVICES INVOLVED: BATCH REVIEW PERFORMED ON 18 FEBRUARY 2022: LINER: VERSAFITCUP DM 01.26.2852MHC DOUBLE MOBILITY HC LINER Ø 52/28 (K092265) LOT 2103099: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 30-APR-2021. EXPIRATION DATE: 2026-04-18. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. STEM: QUADRA-H 01.12.033 CEMENTLESS, HA COATED LAT STEM # 3 (K082792) LOT 182834: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 19-SEP-2018. EXPIRATION DATE: 2023-09-05. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CUP: VERSAFITCUP 01.26.52MB ACETABULAR SHELL Ø 52 (K083116) LOT 188841: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 15-FEB-2019. EXPIRATION DATE: 2024-02-06. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 0

THE PATIENT HAD A PRIMARY HIP SURGERY ON (B)(6) 2019. ON (B)(6) 2021, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN WAS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. PRESENTLY, ON (B)(6) 2022, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REMOVED ALL COMPONENTS, PERFORMED A WASHOUT, AND REIMPLANTED NEW COMPONENTS. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
531031 BALL HEADS: MECTACER 01.29.202 BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 28 SIZE M CERAMIC BALL HEAD LZO MEDACTA INTERNATIONAL SA 01.29.202 2012905 07630030809132

Patients

Seq Age Sex Outcome Treatment
1 Female Required Intervention