FDA Adverse Event Other Summary report: N

ALLEN EASY LOCK BLADE CLAMP

MDR report key: 1381891 · Received May 13, 2009

Report

Report Number
1221538-2009-00004
Event Type
Other
Date Received
May 13, 2009
Date of Event
April 8, 2009
Report Date
May 13, 2009
Manufacturer
ALLEN MEDICAL SYSTEMS
Product Code
HXD
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SALES REP VISITED THE FACILITY IMMEDIATELY AFTER THE INCIDENT. DURING THAT VISIT, THE REP WITNESSED THE USERS INSTALL THE RAIL CLAMPS IN SUCH A WAY THAT DAMAGE TO THE UNITS OR A FAILURE COULD RESULT DURING USE. IN-SERVICE DEVICE INSTRUCTIONS AND CORRECTIVE TRAINING WERE BOTH PROVIDED. ALLEN HAS BEEN NOTIFIED THAT THE DAMAGE CLAMP WILL NOT BE RETURNED FOR OUR EVALUATION BUT MAY BE EVALUATED AT THEIR FACILITY. AT THIS TIME, THERE ARE NO PLANS TO DO AN ENGINEERING EVALUATION IN THE FIELD. THIS INCIDENT IS BEING REPORTED IN RESPONSE TO A MEDWATCH REPORT COMPLETED BY THE USER FACILITY.

Description of Event or Problem · 1

ON (B)(6), 2009, AN ALLEN SALES REP WAS CONTACTED BY A CUSTOMER REPORTING A NON-INJURY INCIDENT INVOLVING AN ALLEN CAMP. A WRITTEN REPORT FROM THE HOSPITAL FOLLOWED AND WAS RECEIVED BY ALLEN ON (B)(6). ACCORDING TO THE REPORT, THE INCIDENT OCCURRED AFTER THE COMPLETION OF THE PROCEDURE, WHEN THE CLAMP CAME FREE FROM THE TABLE RAIL. THE PATIENT WAS STILL ON THE TABLE AT THE TIME BUT NOT INJURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ALLEN EASY LOCK BLADE CLAMP EASY LOCK BLADE CLAMP HXD ALLEN MEDICAL SYSTEMS A-40040-A2 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK