FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG CARD

MDR report key: 13810336 · Received March 18, 2022

Report

Report Number
1221359-2022-01595
Event Type
Malfunction
Date Received
March 18, 2022
Report Date
June 6, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC
Product Code
QKP
UDI-DI
10811877011290
PMA / PMN Number
EUA202537
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. H10: REFERENCE MFR NUMBERS: 1221359-2022-01593, 1221359-2022-01594.

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 179269 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-000/ LOT 179269, TEST BASE PART NUMBER 195-430H/ LOT 172977. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 179269 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO THE SPECIFIC PATIENT/VALIDATION SAMPLES. H10: REFERENCE MFR NUMBERS: 1221359-2022-01593 THROUGH 1221359-2022-01595.

Description of Event or Problem · 0

THE CONSUMER REPORTED 3 FALSE POSITIVE RESULTS WITH THE BINAXNOW COVID 19 ANTIGEN TEST. CONFIRMATION TESTING WAS PERFORMED BY PCR AND GENERATED NEGATIVE RESULTS. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME WAS PROVIDED. THIS IS NUMBER ONE (1) OF THREE (3) REPORTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2222695 BINAXNOW COVID-19 AG CARD LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC 179269 10811877011290

Patients

Seq Age Sex Outcome Treatment
1 Unknown