FDA Adverse Event
Malfunction
Summary report: N
OMNIPOD INSULIN PUMP
MDR report key: 1379399
·
Received March 4, 2009
Report
- Report Number
- 3004464228-2009-00151
- Event Type
- Malfunction
- Date Received
- March 4, 2009
- Date of Event
- February 9, 2009
- Report Date
- February 9, 2009
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE RETURNED PRODUCT EVALUATION CONFIRMED AN INTERNAL LEAK WHICH CAUSED DAMAGE TO THE DEVICE AFTER BEING FILLED WITH INSULIN. THE USER IS INSTRUCTED IN THE USER GUIDE TO FREQUENTLY MONITOR THEIR BG LEVELS. BY FOLLOWING THIS RECOMMENDATION, THE USER WAS AWARE OF HIGH BG LEVELS AND WAS ABLE TO START A NEW POD SUCCESSFULLY. THE LOT WAS FOUND TO HAVE MET ALL ACCEPTANCE CRITERIA.
Description of Event or Problem · 1
CUSTOMER CALLED TO SAY THAT HE WAS HAVING ELEVATED BLOOD GLUCOSE LEVELS THAT RANGED BETWEEN 241-293 MG/DL BEFORE THE POD ALARMED. CUSTOMER WAS ABLE TO ACTIVATE A NEW POD SUCCESSFULLY. NO FURTHER ISSUES WERE IDENTIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L11873 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | Other |