BLOOD GAS ANALYZER
Report
- Report Number
- 3002807968-2005-00014
- Event Type
- Other
- Date Received
- December 8, 2005
- Date of Event
- October 23, 2005
- Report Date
- December 5, 2005
- Manufacturer
- RADIOMETER MEDICAL APS
- Product Code
- CHL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THE SYMPTOMS ARE CONSISTENT WITH A COAGULATED BLOOD CLOT OBSTRUCTING THE PASSAGE OF THE BLOOD SAMPLE INTO THE ANALYZER - FROM A NEONATE. THIS BLOOD CLOT IN THE CAPILLARY MAY HAVE CAUSED THAT THE SAMPLE WERE NOT ASPIRATED INTO THE ANALYZER AT FIRST. BUT A VACUUM WERE CREATED IN THE ABL 735. A SUDDEN RELEASE OF THE CLOT MAY HAVE CAUSED A POOR LIQUID TRANSPORT ON THE ABL OF THE SAMPLE. ALSO IT MAY HAVE CAUSED A DILUTION OF THE SAMPLE. IT HAS NOT BEEN POSSIBLE TO GET ALL THE RELEVANT INFO FOR MAKING A CONCLUSION TO THIS CASE YET. THEREFORE, A FINAL REPORT WILL BE SEND WITHIN A MONTH.
A NEONATE WAS INITIALLY TESTED TO HAVE A TOO HIGH BILIRUBIN VALUE (300 MMOL/L). WHEN THE NEONATE WAS TESTED AGAIN THE RESULT THIS TIME WAS 76 MMOL/L. THE DOCTOR IN CHARGE DECIDED THAT THE BILIRUBIN VALUE WAS TOO LOW COMPARED TO THE EARLIER VALUE. IN THE AFTERNOON THE SAME DAY A NEW SAMPLE FROM THE SAME CHILD WAS TESTED - THIS TIME, IT SHOWED THE VALUE 288 MMOL/L. THE DOCTOR TRUSTED THIS RESULT. WHEN THE ABL ANALYZER TESTED THE SAMPLE WITH THE TOO LOW RESULT, THE ABL 735 ERROR MESSAGES WERE GIVEN. FURTHERMORE QUESTION MARKS TO ALL PARAMETERS MEASURED IN THE PH/BG MODULE, AND TO PARAMETERS DEPENDENT ON THE REFERENCE ELECTRODE WERE GIVEN. IT IS BELIEVED THAT THE POOR SAMPLE TRANSPORT DID ALSO GIVE ERRONEOUS RESULTS FOR THE REST OF THE PARAMETERS MEASURED IN THE EL/MET MODULE AND OXI MODULE - AND THESE PARAMETERS WERE GIVEN WITHOUT QUESTION MARKS. IF THE LOW RESULT HAD BEEN TRUSTED, THE WORST CASE SCENARIO IS THAT THE CHILD COULD HAVE HAD A BRAIN DAMAGE - BUT THE LOW BILIRUBIN VALUE WAS INCONSISTENT WITH THE CLINICIANS EXPECTED PICTURE OF A POSSIBLE DECREASE IN BILIRUBIN OF THE NEONATE. NO OTHER DEVICE WAS INVOLVED IN THE INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BLOOD GAS ANALYZER | ABL700/ABL800 SERIES | CHL | RADIOMETER MEDICAL APS | ABL 735 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4 DA | Other |