ENDURITY PACEMAKER, DR, MRI RESTRICTED
Report
- Report Number
- 2017865-2022-04997
- Event Type
- Injury
- Date Received
- March 17, 2022
- Date of Event
- January 24, 2022
- Report Date
- January 4, 2023
- Manufacturer
- ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
- Product Code
- LWP
- UDI-DI
- 05414734509589
- PMA / PMN Number
- P140033
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
CORRECTION: D4 - PREVIOUSLY NOTED MODEL NUMBER, SERIAL NUMBER, LOT NUMBER, EXPIRATION DATE AND UDI SHOULD BE PM2272, 5030208, (B)(6), 30 SEP 2019, AND (B)(4), RESPECTIVELY. G3 - PREVIOUSLY NOTED PMA SHOULD BE P140033. H4 - PREVIOUSLY NOTED DEVICE MANUFACTURE DATE SHOULD BE 13 MAR 2018.
A DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED AND ALL REQUIRED MANUFACTURING PROCESSES AND INSPECTIONS STEPS WERE CONFIRMED TO BE COMPLETED PER THE REQUIREMENTS. THE DEVICE MET SPECIFICATIONS PRIOR TO LEAVING ABBOTT MANUFACTURING FACILITIES.
THE REPORTED EVENT OF PREMATURE BATTERY DEPLETION WAS NOT CONFIRMED. INTERROGATION OF THE DEVICE REVEALED THE DEVICE WAS AT END OF LIFE (EOL). ANALYSIS OF THE DEVICE IMAGE AND A LONGEVITY CALCULATION WAS PERFORMED AND FOUND THE BATTERY DEPLETION WAS NORMAL BASED ON THE DEVICE USAGE. BASED ON THIS INFORMATION, THERE WAS NO EVIDENCE OF PREMATURE BATTERY DEPLETION.
IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A FOLLOW-UP IN CLINIC. UPON INTERROGATION, IT WAS NOTED THAT THE PACEMAKER EXHIBITED PREMATURE BATTERY DEPLETION. THE PACEMAKER WAS EXPLANTED AND REPLACED. THE PATIENT WAS IN STABLE CONDITION THROUGHOUT THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1246333 | ENDURITY PACEMAKER, DR, MRI RESTRICTED | IMPLANTABLE PACEMAKER PULSE GENERATOR | LWP | ST. JUDE MEDICAL, INC.(CRM-SYLMAR) | PM2272 | P000058450 | 05414734509589 |
| 2301302 | ENDURITY PACEMAKER, DR, MRI RESTRICTED | IMPLANTABLE PACEMAKER PULSE GENERATOR | LWP | ST. JUDE MEDICAL, INC.(CRM-SYLMAR) | PM2272 | P000058450 | 05414734509589 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |