FDA Adverse Event Malfunction Summary report: N

BD BBL¿ CDC ANAEROBE 5% SHEEP BLOOD AGAR

MDR report key: 13792881 · Received March 17, 2022

Report

Report Number
1119779-2022-00438
Event Type
Malfunction
Date Received
March 17, 2022
Date of Event
February 18, 2022
Report Date
March 14, 2022
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
JSG
PMA / PMN Number
K803025
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: 252467, LOT# 1342900. THE COMPLAINT WAS NOT CONFIRMED AS THE REPORTED DEFECT ON A PICTURE. WE CONFIRMED NO ISSUE IN PHOTO. WE ONLY CONFIRMED WATER SPOT MARK ON THE AGAR SURFACE. THIS PHENOMENON IS IN SPEC QC ATTRIBUTE TEST. WE CANNOT CONTROL THE MOISTURE DURING TRANSPORTATION. WE HAVEN'T HAD SAME REPORT FROM OTHER CUSTOMER. NO ISSUE IN DHR. NO ISSUE IN RETENTION SAMPLE. WE WILL CONTINUE TO MONITOR THIS ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD BBL¿ CDC ANAEROBE 5% SHEEP BLOOD AGAR THE CUSTOMER OBSERVED BIOLOGICAL CONTAMINATION AND WATER MARKS. THIS EVENT OCCURRED 5 TIME(S). THERE IS NO INDICATION THAT FALSE RESULTS WERE REPORTED TO HEALTH CARE PRACTITIONERS. THERE IS NO INDICATION OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THERE ARE SOME BACTERIA ON THE PLATE WITH WATER MARKS 1-3, BUT NONE OF THEM GROW. JUST A BUNCH OF WATER MARKS."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
766047 BD BBL¿ CDC ANAEROBE 5% SHEEP BLOOD AGAR CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL JSG BECTON, DICKINSON & CO. (SPARKS) 1342900

Patients

Seq Age Sex Outcome Treatment
1 Unknown