MENTOR MEMORYGEL BREAST IMPLANT
Report
- Report Number
- 1645337-2022-02877
- Event Type
- Injury
- Date Received
- March 17, 2022
- Date of Event
- January 20, 2022
- Manufacturer
- MENTOR TEXAS
- Product Code
- FTR
- UDI-DI
- 00081317000549
- PMA / PMN Number
- P030053
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ON MARCH 17, 2022, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON APRIL 13, 2022, MENTOR COMPLETED EVALUATION OF THE DEVICE. DEVICE EVALUATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE SMOOTH HPG, 550CC BREAST IMPLANT WAS FOUND TO BE RUPTURED. ADDITIONALLY, A CREASE/FOLD WAS NOTED ON THE EDGE OF THE TEAR. THE EVALUATION DETERMINED THAT THE CAUSE OF THE RUPTURE IS CONSISTENT WITH NORMAL WEAR. THE INSTRUCTIONS FOR USE STATE THAT RUPTURES CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT IT IS MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. THE FOLLOWING MAY CAUSE IMPLANTS TO RUPTURE: STRESSING THE IMPLANT DURING IMPLANTATION OR OTHER PROCEDURES AND WEAKENING IT; FOLDING OR WRINKLING OF THE IMPLANT SHELL. BREAST IMPLANTS MAY ALSO WEAR OVER TIME. MOST WOMEN UNDERGOING AUGMENTATION OR RECONSTRUCTION WITH A MAMMARY PROSTHESIS WILL EXPERIENCE SOME PAIN POSTOPERATIVELY. WHILE PAIN NORMALLY SUBSIDES IN MOST WOMEN AS THEY HEAL FROM SURGERY, IT CAN BECOME A CHRONIC PROBLEM IN OTHER WOMEN. CHRONIC PAIN CAN BE ASSOCIATED WITH A VARIETY OF FACTORS. SURGEONS SHOULD INSTRUCT THEIR PATIENTS TO INFORM THEM IF THERE IS SIGNIFICANT PAIN OR IF PAIN PERSISTS. PAIN IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. A SECOND PRODUCT WAS RECEIVED (LOT-5993166). NO ADVERSE EVENTS WERE REPORTED FOR THIS CONCOMITANT (CONTRALATERAL) DEVICE, THEREFORE NO FURTHER ANALYSIS IS REQUIRED. IT SHOULD BE NOTED THAT AS PART OF THE MENTOR¿S QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RUPTURE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A (B)(6) CAUCASIAN FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH A 500CC MENTOR MEMORYGEL BREAST IMPLANT AND EXPERIENCED BREAST PAIN AND A RUPTURE ON THE LEFT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED VIA A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2022.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 937237 | MENTOR MEMORYGEL BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | MENTOR TEXAS | 3505504BC | 5993166 | 00081317000549 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Female | Required Intervention |