BD SEDI-40
Report
- Report Number
- 2243072-2022-00362
- Event Type
- Malfunction
- Date Received
- March 17, 2022
- Date of Event
- March 7, 2022
- Report Date
- June 23, 2022
- Manufacturer
- BECTON DICKINSON
- Product Code
- JKA
- PMA / PMN Number
- K953994
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
IN THIS MDR, BD CORPORATE HEADQUARTERS IN (B)(4) HAS BEEN LISTED AS (B)(4) IS AN OEM MANUFACTURING SITE. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN.
H6: INVESTIGATION SUMMARY: INSTRUMENT SEDI 40 19-42033 WAS RETURNED TO THE MANUFACTURER FOR SERVICE WITH RESPECT TO THE REPORTED DEFECT ¿ DAMAGED POWER SUPPLY. THE INSTRUMENT WAS EVALUATED BY VISUAL EXAMINATION AND FUNCTIONAL TESTING AND IT WAS FOUND THAT POWER SUPPLY PCB WAS DEFECTIVE. THE POWER SUPPLY PCB WAS REPLACED. AFTER REPAIR THE INSTRUMENT PASSED ALL FURTHER QUALITY CHECKS. THIS COMPLAINT HAS BEEN CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED WHEN USING THE BD SEDI-40 THERE WAS A BROKEN/DAMAGED INSTRUMENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "THE POWER SUPPLY INLET WAS DESTROYED."
IT WAS REPORTED WHEN USING THE BD SEDI-40 THERE WAS A BROKEN/DAMAGED INSTRUMENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "THE POWER SUPPLY INLET WAS DESTROYED."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1246288 | BD SEDI-40 | ESR INSTRUMENT | JKA | BECTON DICKINSON | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |