CART 9734056 S7 STAFF SHRT 100-120V INTL
Report
- Report Number
- 1723170-2022-00377
- Event Type
- Malfunction
- Date Received
- March 15, 2022
- Date of Event
- March 11, 2022
- Report Date
- April 13, 2022
- Manufacturer
- MEDTRONIC NAVIGATION, INC
- Product Code
- HAW
- PMA / PMN Number
- K050438
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
CONCOMITANT MEDICAL PRODUCTS: OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 9735339R, SERIAL/LOT #: UNKNOWN, PRODUCT ID: 9733438, SERIAL/LOT #: UNKNOWN, PRODUCT ID: 9735225R, SERIAL/LOT #: UNKNOWN, PRODUCT ID: 9733582, SERIAL/LOT #: UNKNOWN, PRODUCT ID: 9733575, SERIAL/LOT #: UNKNOWN. A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. THE REPORTED ISSUE WAS CONFIRMED. THE SYSTEM POSITION SENSOR UNIT (PSU), SENSOR CONTROL UNIT (SCU), COMPUTER, INTERNAL SCU TO PSU CABLE, AND EXTERNAL SCU TO PSU CABLE WERE REPLACED. THE SYSTEM THEN PASSED THE SYSTEM CHECKOUT AND WAS FOUND TO BE FULLY FUNCTIONAL. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
H3 - EVALUATION OF THE RETURNED SYSTEM POSITION SENSOR UNIT (PSU), SENSOR CONTROL UNIT (SCU), COMPUTER, INTERNAL SCU TO PSU CABLE, AND EXTERNAL SCU TO PSU CABLE FOUND NO FAULT WITH THE PRODUCTS. H6 - EVALUATION CODES C19 AND D14 APPLY TO THE SYSTEM POSITION SENSOR UNIT (PSU), SENSOR CONTROL UNIT (SCU), COMPUTER, INTERNAL SCU TO PSU CABLE, AND EXTERNAL SCU TO PSU CABLE. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. DURING AN INSPECTION, A BEEP WAS HEARD FROM THE SYSTEM POWER SUPPLY UNIT. THE FAULT "X" WAS ALSO OBSERVED FOR THE CAMERA AND INSTRUMENT ICONS. THE "NOT CONNECTED" MESSAGE WAS ALSO DISPLAYED ON THE SYSTEM. THE SYSTEM RECOVERED AFTER WAITING A FEW SECONDS, BUT IT WOULD REAPPEAR AFTER A WHILE. THE PARTS WOULD BE REPLACED AT A LATER DATE. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 695434 | CART 9734056 S7 STAFF SHRT 100-120V INTL | NEUROLOGICAL STEREOTAXIC INSTRUMENT | HAW | MEDTRONIC NAVIGATION, INC | 9734056 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | "SEE H10...." |