IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM
Report
- Report Number
- 0002023141-2022-00637
- Event Type
- Injury
- Date Received
- March 15, 2022
- Date of Event
- February 23, 2022
- Report Date
- August 18, 2022
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024020047
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). PATIENT IDENTIFIER UNKNOWN / NOT PROVIDED, AGE AND DATE OF BIRTH UNKNOWN / NOT PROVIDED, PATIENT SEX UNKNOWN / NOT PROVIDED, WEIGHT UNKNOWN / NOT PROVIDED, FAX NUMBER UNKNOWN / NOT PROVIDED,AND ADDITIONAL PMA/510(K) NUMBER ¿ K011028/K013227. CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: H6: INVESTIGATION TYPE CODES WERE ADDED: 3331, 4109 AND 4111. H6: INVESTIGATION FINDINGS CODE WAS ADDED: 3252. H6: INVESTIGATION CONCLUSIONS CODES WERE ADDED: 4307. ONE (1) IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM (TSVWB10) WAS RETURNED FOR INVESTIGATION. VISUAL EVALUATION OF THE AS RETURNED PRODUCT IDENTIFIED THE IMPLANT FRACTURED AT THE COLLAR REGION. SIGNS OF USE. BASED ON THE EVALUATION, THE DEVICE MALFUNCTION HAS OCCURRED. HOWEVER, THERE IS NO EXISTING NONCONFORMANCE/CAPA/HHE/D/IE/PRODUCT HOLDS AGAINST THE REPORTED DEVICE THAT COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. ZIMMER BIOMET QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT. THEREFORE, BASED ON THE AVAILABLE INFORMATION, THE PRODUCT WAS LIKELY WITHIN SPECIFICATIONS AND LIKELY CONFORMING WHEN IT LEFT ZIMMER BIOMET. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1251414). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1251414) FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. THE REPORTED EVENT COULD NOT BE RECREATED DUE TO THE NATURE OF THE DENTAL DEVICE AND EVENT (FRACTURE) AND THE COMPLAINT IS RELATED TO THE FUNCTIONAL PERFORMANCE OF THE DEVICE. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. HOWEVER, BASED ON THE INVESTIGATION AND RISK FILE REVIEW, THE MOST LIKELY CAUSES DETERMINED FROM THE INVESTIGATION ARE IMPLANT NOT PLACED PROPERLY; CLINICIAN ERROR. NO FURTHER INVESTIGATION AND NO IMMEDIATE CAPA/HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #19 FRACTURED DURING INSTALLATION AND WAS REMOVED.
NO ADDITIONAL EVENT INFORMATION AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 47471 | IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | TSVWB10 | 1251414 | 00889024020047 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |