ENDOPATH ETS
Report
- Report Number
- 1527736-1997-03381
- Event Type
- Malfunction
- Date Received
- December 11, 1997
- Date of Event
- November 12, 1997
- Report Date
- November 12, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- KOG
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
DEVICE RETURNED WITH NO LOT ID. THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION. THE RESULTS OF THE INVESTIGATION CONDUCTED BY THE APPROPRIATE ENGINEERS AND TECHNICIANS ARE LISTED BELOW. VISUAL INSPECTIONS & RESULTS: ANY OTHER NOTICEABLE DAMAGE, NO; BATCH NUMBER, K00200; CARTRIDGE PAN IN PLACE/CONDITION, YES/GOOD; CONDITION OF DRIVERS, GOOD; LOCKOUT TABS ON PAN CONDITION, FIRED AND POSITION/CONDITION OF WEDGE SLEDS, FULLY FIRED. FUNCTIONAL TESTS & RESULTS: CONDITION OF CLAMP FIRST LOCKOUT, N/A; CONDITION OF CLAMPING MECHANISM, N/A; CONDITION OF FIRING MECHANISM, N/A; CONDITION OF KNIFE, N/A; CONDITION OF WEDGE BANDS, N/A; IS HYPER LOCKOUT CONDITION PRESENT, N/A AND RESULT OF ATTEMPTED FIRING, N/A. ANALYSIS CONCLUSION: BASED UPON THE INQUIRY INFO REC'D, THE VISUAL EXAMINATION, AND THE FUNCTIONAL TESTING, NO CONCLUSION COULD BE REACHED AS TO WHY THE CARTRIDGE REPORTEDLY PRODUCED "MALFORMED STAPLES" DURING SURGERY. THE CARTRIDGE WAS RETURNED FULLY FIRED AND IN GOOD PHYSICAL CONDITION. NO TESTING COULD BE PERFORMED AS THE CARTRIDGE HAD BEEN FULLY FIRED. EACH INSTRUMENT IS EVALUATED DURING THE ASSEMBLY PROCESS TO ENSURE IT FUNCTIONS PROPERLY.
IT WAS REPORTED DURING A LAPAROSCOPIC OOPHORECTOMY THE SURGEON HAD FIRED THE DEVICE AND ONLY SOME STAPLES FORMED AND THE DEVICE DID NOT COMPLETELY CUT THE TISSUE. THERE WAS NO FURTHER INFO SURROUNDING THIS DEVICE. THE REP WAS CALLED INTO THE ROOM. A SECOND ETS WAS OPENED, THE SURGEON ATTEMPTED TO FIRE THE INSTRUMENT AND THE CARTRIDGE WOULD NOT FIRE. A NEW CARTRIDGE WAS PLACED ON THE ETS AND IT FIRED PROPERLY. THERE WAS NO PT CONSEQUENCE. 11/21/97 THE REP REPORTS THE FIRST ETS WAS NOT SAVED AND IT IS UNKNOWN IF IT WAS THE INITIAL OR THE SECOND FIRING OF IT. REP WAS PRESENT DURING THE FIRING OF THE SECOND ETS. THE INITIAL FIRING WAS FINE. ON THE SECOND RELOAD IT WOULD NOT FIRE. THE REP NOTED DRIVERS IN THE END OF THE CARTRIDGE. THE SAME DEVICE WAS RELOADED AND THE NEXT TWO FIRINGS WORKED PROPERLY. THERE WAS NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH ETS | LINEAR CUTTER | KOG | ETHICON ENDO-SURGERY, INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |