FDA Adverse Event Malfunction Summary report: N

SUPERA BILIARY STENT DELIVERY SYSTEM

MDR report key: 1375283 · Received February 20, 2009

Report

Report Number
3005325609-2009-00002
Event Type
Malfunction
Date Received
February 20, 2009
Date of Event
January 22, 2009
Report Date
January 22, 2009
Manufacturer
IDEV TECHNOLOGIES, INC.
Product Code
FGE
PMA / PMN Number
K071646
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE DEVICES WERE DISCARDED AT THE HOSPITAL. THE DEVICE HISTORY RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED WITH THE DEVICE'S LOT OR MANUFACTURING PROCESSES. THIS EVENT WAS THE FIRST REPORTED EVENT WHERE A MARKER BAND DISLODGED AND IT REMAINED IN THE PATIENT. PREVIOUS INVESTIGATIONS REGARDING THIS TYPE OF EVENT HAVE CONCLUDED THAT THE REPORTED MARKER BAND MOVEMENT/DISLODGEMENT EVENTS ARE RELATED TO THE USE OF ONE TYPE OF INTRODUCER SHEATH (SAME MANUFACTURER AND SAME BRAND NAME) WITH THE STENT DELIVERY CATHETER. WE HAVE RECEIVED TWO OF THE INTRODUCER SHEATH'S BACK FOR ANALYSIS AND FOUND THAT WHEN THIER VALVE IS TIGHTENED, A PIN GAUGE TEST RESULTS IN LESS THAN 7FR. THIS TIGHTNESS RESULTS IN THE SHEATH CATCHING THE MARKER BAND CAUSING IT TO MOVE/DISLODGE. THE MANUFACTURE OF THE SHEATH HAS BEEN INFORMED OF THE EVENTS. THERE HAVE BEEN NO SERIOUS INJURIES OR DEATHS RELATED TO THESE EVENTS.

Description of Event or Problem · 1

THE REPORTER INDICATED THAT THE HUB ON THE INTRODUCER SHEATH WAS TOO TIGHT AND CAUSED ONE OF THE MARKER BANDS TO COME OFF OF THE STENT DELIVERY CATHETER. IT WAS REPORTED THAT AN ATTEMPT WAS MADE TO RETRIEVE THE MARKER BAND. THE PHYSICIAN DEVIDED THAT SINCE THE MARKER BAND HAD LODGED IN A DEAD END VESSEL OFF THE PROFUNDA THAT IT WOULD HAVE NO LONG-TERM CLINICAL RELEVANCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SUPERA BILIARY STENT DELIVERY SYSTEM FGE IDEV TECHNOLOGIES, INC. S-05-120-120-G2 F00253038

Patients

Seq Age Sex Outcome Treatment
1 56 YR SHEATH: 7F PINNACLE