FDA Adverse Event Death Summary report: N

PIPELINE

MDR report key: 13748478 · Received March 12, 2022

Report

Report Number
2029214-2022-00403
Event Type
Death
Date Received
March 12, 2022
Date of Event
December 27, 2021
Report Date
March 11, 2022
Manufacturer
MICRO THERAPEUTICS, INC. DBA EV3
Product Code
OUT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

HUIJIAN GE, XIHENG CHEN, KAI LIU, YANG ZHAO, LONGHUI ZHANG, PENG LIU, YUHUA JIANG, HONGWEI HE, MING LV, YOUXIANG LI; FRONTIERS IN NEUROLOGY; 2022; 13 (843839); ENDOVASCULAR TREATMENT OF LARGE OR GIANT BASILAR ARTERY ANEURYSMS USING THE PIPELINE EMBOLIZATION DEVICE: COMPLICATIONS AND OUTCOMES; DOI:10.3389/FNEUR.2022.843839. MEDTRONIC RECEIVED INFORMATION IN A LITERATURE ARTICLE OF PATIENT'S HAVING COMPLICATIONS WITH PIPELINE EMBOLIZATION DEVICES (PED)S. 29 PATIENTS (18 MEN, 11 WOMEN) WITH LARGE OR GIANT BASIL ARTERY (BA) ANEURYSMS WERE RETROSPECTIVELY EXAMINED. MEAN AGE WAS 44.1. THE STUDY DESCRIBES THE EXPERIENCE USING THE PED TO TREAT THESE ANEURYSMS. RETROSPECTIVELY COLLECTED THE DATA OF CONSECUTIVE PATIENTS WITH LARGE OR GIANT BA ANEURYSMS WHO WERE ELECTIVELY TREATED USING THE PED AT OUR CENTER FROM JANUARY 2016 TO OCTOBER 2020. ALL PATIENTS RECEIVED DUAL ANTIPLATELET THERAPY (CLOPIDOGREL 75 MG/D AND ASPIRIN 100 MG/D) FOR AT LEAST 5 DAYS BEFORE THE PROCEDURE. A TRIAXIAL GUIDE-CATHETER SYSTEM USING A 6-FR COOK CATHETER (COOK MEDICAL, BLOOMINGTON, IN, USA), 5-FR OR 6-FR NAVIEN GUIDING CATHETER (MEDTRONIC), AND MARKSMAN MICROCATHETER WAS USED TO DEPLOY THE PED. PERIPROCEDURAL COMPLICATIONS WERE DEFINED AS THOSE THAT DEVELOPED WITHIN 30 DAYS OF THE PROCEDURE. FAVORABLE OUTCOME WAS DEFINED AS MRS SCORE LESS THAN OR EQUAL TO 2; POOR OUTCOME WAS DEFINED AS SCORE GREATER THAN 2. SEVENTEEN BA ANEURYSMS WERE LARGE AND 12 WERE GIANT. MEAN ANEURYSM SIZE WAS 22.2. FUSIFORM ANEURYSMS WERE MORE COMMON (18 PATIENTS, 62.1%). FIFTEEN BA ANEURYSMS (51.7%) WERE CLASSIFIED AS DISTAL AND 14 AS PROXIMAL. EIGHT ANEURYSMS (25.8%) PRESENTED WITH INTRAANEURYSMAL THROMBUS; TWO (CASES 12 AND 16) HAD BEEN PREVIOUSLY COILED AND REQUIRED RETREATMENT BECAUSE OF RECANALIZATION. FORTY-ONE PEDS WERE IMPLANTED IN THE 29 PATIENTS. ADJUNCTIVE COILING OR BALLOON PLACEMENT WAS PERFORMED IN 12 PATIENTS (41.4%). SUFFICIENT PED OPENING WAS ACHIEVED IN ALL PATIENTS. COMPLETE OCCLUSION WAS ACHIEVED IN 17 ANEURYSMS (74%) AND NEAR-COMPLETE OCCLUSION IN THREE (13%). PERIPROCEDURAL COMPLICATIONS INCLUDED ISCHEMIC STROKE IN FIVE PATIENTS, WORSENING MASS EFFECT IN THREE, AND DELAYED CEREBRAL HEMORRHAGE IN FOUR. AMONG THE ISCHEMIC STROKES, FOUR WERE PERFORATOR STROKES AND ONE WAS EMBOLIC. ONE CASE EXPERIENCED ACUTE ONSET DYSARTHRIA AND RIGHT HEMIPLEGIA 12 HOURS AFTER PLACEMENT OF FOUR PEDS. CT AND MRI SHOWED A LARGE BRAINSTEM INFARCT WITH NO HEMORRHAGE. AFTER A 24-H TIROFIBAN INFUSION, THE PATIENT¿S SYMPTOMS GRADUALLY RESOLVED. HOWEVER, ON POSTPROCEDURE DAY 3, HEADACHE AND VOMITING DEVELOPED, WHICH RAPIDLY PROGRESSED TO LOSS OF CONSCIOUSNESS, CARDIORESPIRATORY ARREST, AND DEATH. DELAYED ANEURYSMAL RUPTURE WAS THE PRESUMED CAUSE OF DEATH. OTHER CASES RECOVERED COMPLETELY BUT 2 CASES HAD MILD RESIDUAL SINGLE-LIMB PARESIS. THREE PATIENTS WITH GIANT ANEURYSMS EXPERIENCED WORSENED MASS EFFECT AFTER TREATMENT; ALL PRESENTED WITH INITIAL SYMPTOMS OF BRAINSTEM COMPRESSION. 2 CASES EXPERIENCED ABDUCENS NERVE PALSY AFTER THE PROCEDURE; HOWEVER, MRI DID NOT SHOW AN INFARCTION. ANOTHER CASE DEVELOPED DYSPNEA BECAUSE OF ANEURYSMAL BRAINSTEM COMPRESSION AFTER TREATMENT AND MAINTAINED A TRACHEOTOMY UNTIL HE DIED OF SEVERE PULMONARY INFECTION 11MONTHS LATER. IN THE PERIOPERATIVE PERIOD, FOUR PATIENTS WITH ANEURYSM SIZE RANGING FROM 22 TO 40MM DEVELOPED FATAL SUBARACHNOID HEMORRHAGE. IN-STENT THROMBOSIS OCCURRED IN FOUR PATIENTS. ANGIOGRAPHY AFTER THROMBECTOMY REVEALED THAT THESE ANEURYSMS WERE NOT COMPLETELY OCCLUDED. AT LAST FOLLOW-UP, 2 OF THOSE CASES HAD MRS SCORES OF 4 AND 3, WHILE CASES THE OTHER 2 DIED A SHORT TIME AFTER THROMBECTOMY. TO SUMMERIZE TWENTY ONE PATIENTS (72.4%) ACHIEVED A FAVORABLE CLINICAL OUTCOMES. NINE PATIENTS (31%) EXPERIENCED A POOR CLINICAL OUTCOMES, INCLUDING TWO PATIENTS WITH SEVERE DISABILITY AND SEVEN PATIENTS WHO DIED. OVERALL RATES OF MORBIDITY AND MORTALITY WERE 10.3 AND 24.2%, RESPECTIVELY. THE MAIN CAUSES OF DEATH WERE DELAYED ANEURYSMAL RUPTURE (N = 4), IN-STENT THROMBOSIS (N=2) AND WORSENED MASS EFFECT (N=1).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2113668 PIPELINE INTRACRANIAL ANEURYSM FLOW DIVERTER OUT MICRO THERAPEUTICS, INC. DBA EV3 NV UNK PIPELINE UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 44 YR Male Death