FDA Adverse Event Malfunction Summary report: N

COOK COPE-LOOP NEPHROSTOMY SET

MDR report key: 13737845 · Received March 11, 2022

Report

Report Number
1820334-2022-00386
Event Type
Malfunction
Date Received
March 11, 2022
Date of Event
December 1, 1994
Report Date
November 16, 2022
Manufacturer
COOK INC
Product Code
LJE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

COMMON NAME & PRODUCT CODE = KOB, CATHETER, SUPRAPUBIC (AND ACCESSORIES). PMA/510(K) NUMBER = UNAVAILABLE AS THE DEVICE LOT NUMBER, RPN, AND GPN ARE UNKNOWN. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNCHANGED, UNKNOWN OR UNAVAILABLE. EVENT DESCRIPTION: AS REPORTED IN THE LITERATURE ARTICLE, "UROLOGIST DIRECTED PERCUTANEOUS NEPHROSTOMY TUBE PLACEMENT," A COOK COPE-LOOP NEPHROSTOMY SET BECAME OCCLUDED AFTER PLACEMENT. THE AUTHORS REPORT THE FOLLOWING: "A RETROSPECTIVE REVIEW OF 100 CONSECUTIVELY PLACED NEPHROSTOMY TUBES IN 18 PATIENTS (53 WOMEN AND 29 WOMEN) WAS DONE. ALL NEPHROSTOMY TUBES WERE PLACED BY HOUSE STAFF UNDER THE SUPERVISION OF A SINGLE ATTENDING UROLOGIST. VARIED NEPHROSTOMY TUBES WERE USED TO GAIN EXPERIENCE WITH A RANGE OF DRAINAGE SYSTEMS. "PATIENTS WERE PLACED IN THE PRONE POSITION WITH THE KNEES FLEXED AND ALL PRESSURE POINTS PADDED FOR NEPHROSTOMY TUBE INSERTION. PERCUTANEOUS ACCESS WAS ROUTINELY OBTAINED WITH A 17.5 GAUGE ANGIOMED 3-PART INITIAL PUNCTURE NEEDLE UNDER ULTRASONIC GUIDANCE WITH A 3.5 MHZ TRANSDUCER. PRELIMINARY ULTRASONIC IMAGING IDENTIFIED ADJACENT BOWEL AND OTHER INTERNAL ORGANS. KNOWLEDGE OF THE APPROXIMATE DISTANCE FROM THE SKIN TO THE COLLECTING SYSTEM HELPED ESTIMATE NEEDLE PUNCTURE DEPTH AND, IN ADDITION, MALROTATED KIDNEYS AND RENAL HILAR VESSELS WERE APPRECIATED. ULTRASONIC VISUALIZATION OF THE PUNCTURE NEEDLE WAS MAINTAINED DURING INSERTION. APPROPRIATE PLACEMENT OF THE TIP OF THE PUNCTURE NEEDLE ALSO WAS CONFIRMED BY THE EFFLUX OF URINE AND/OR INJECTION OF RADIOCONTRAST MATERIAL INTO THE COLLECTING SYSTEM. "A FLOPPY J-TIPPED GUIDE WIRE WAS PLACED INTO THE COLLECTING SYSTEM AND DOWN THE URETER WHEN POSSIBLE. FASCIAL DILATION DEVELOPED AN APPROPRIATELY SIZED TRACT. THE NEPHROSTOMY TUBE THEN WAS PLACED OVER THE GUIDE WIRE INTO THE COLLECTING SYSTEM. CONTRAST MEDIUM WAS INJECTED UNDER FLUOROSCOPIC IMAGING TO DEFINE THE ANATOMY, CONFIRM PROPER NEPHROSTOMY TUBE PLACEMENT AND EXCLUDE ANY ASSOCIATED INJURIES. ALL TUBES WERE SECURED DIRECTLY TO THE SKIN WITH 2-ZERO NYLON SUTURES AND SIMPLE 4X4 GAUZE DRESSINGS. AN ADDITIONAL FIXATION POINT WITH TAPE WAS PLACED APPROXIMATELY 20CM. FROM THE INSERTION SITE TO HELP PREVENT TWISTING AND KINKING OF THE TUBE BEFORE CONNECTION TO A CLOSED GRAVITY COLLECTION DEVICE. WHEN COPE TUBES WERE USED THE TIP WAS COILED IN THE RENAL PELVIS WITH THE PULL-STRING. BALLOON RETENTION CATHETERS WERE INFLATED WITH DILUTE CONTRAST MEDIUM TO CONFIRM POSITION. IMMEDIATE AND DELAYED COMPLICATIONS ASSOCIATED WITH NEPHROSTOMY TUBE PLACEMENT WERE IDENTIFIED, AND A DETAILED CLINICAL ASSESSMENT WAS DONE." ¿NEPHROSTOMY TUBES (100) WERE PLACED IN 29 WOMEN AND 53 MEN. A RANGE OF DIFFERENT NEPHROSTOMY TUBE ALTERNATIVES WAS USED DURING THE REVIEW PERIOD, REFLECTING A DESIRE ON OUR PART TO IDENTIFY SUPERIOR METHODS OF NEPHROSTOMY ACCESS TO LIMIT PROCEDURE RELATED AND DELAYED DRAINAGE COMPLICATIONS. WE USED 47 SINGLE J, 28 COOK-COPE, 11 RUSH NEPHROSTOMY, 9 FOLEY URINARY, 3 MALECOT AND 1 RUBBER ROBINSON URINARY CATHETERS AND 1 ARGYLE CHEST TUBE.¿ "COMPLICATIONS RELATED TO THE NEPHROSTOMY TUBES WERE ASSESSED AS BEING EITHER EARLY (INVOLVED IN CATHETER PLACEMENT AND BECOMING CLINICALLY APPARENT WITHIN 24 HOURS) OR LATE (BECOMING APPARENT MORE THAN 24 HOURS AFTER INSERTION). A TOTAL OF 27 COMPLICATIONS WAS NOTED IN 14 PATIENTS." "NEPHROSTOMY TUBE OCCLUSION OCCURRED IN 3, 10F SINGLE J CATHETERS, 1 10F COPE TUBE AND 2 FOLEY CATHETERS (16F AND 18F). OCCLUSION APPEARED TO BE DIRECTLY RELATED TO THE VISCOSITY OF DRAINAGE, WITH PURULENT MATERIAL BEING PRESENT IN 5 OF THE 6 CASES." CITED: MAHAFFEY, K.G., BOLTON, D.M., STOLLER, M.L. (1994). UROLOGIST DIRECTED PERCUTANEOUS NEPHROSTOMY TUBE PLACEMENT. THE JOURNAL OF UROLOGY, 152, 1973-1976. HTTPS://DOI.ORG/10.1016/S0022-5347(17)32282-6 INVESTIGATION - EVALUATION REVIEWS OF INSTRUCTIONS FOR USE (IFU) AND QUALITY CONTROL PROCEDURES WERE CONDUCTED DURING THE INVESTIGATION. THE COMPLAINT DEVICE WAS NOT RETURNED; THEREFORE, NO PHYSICAL EXAMINATIONS COULD BE PERFORMED; HOWEVER, A DOCUMENT BASED INVESTIGATION EVALUATION WAS PERFORMED. THERE IS NO EVIDENCE TO SUGGEST THE PRODUCT WAS NOT MADE TO SPECIFICATIONS. THE CURRENT QUALITY CONTROL PROCEDURES WERE INSPECTED INSTEAD OF THE CONTROLS AT THE TIME OF THE JOURNAL ARTICLE FOR THE SAKE OF RELEVANCE. IT WAS CONCLUDED THAT SUFFICIENT INSPECTION ACTIVITIES ARE IN PLACE TO IDENTIFY DEVICE FUNCTIONALITY PRIOR TO DISTRIBUTION. A REVIEW OF THE DEVICE HISTORY RECORD COULD NOT BE COMPLETED DUE TO LACK OF LOT INFORMATION FROM THE USER FACILITY. A REVIEW OF COMPLAINT HISTORY RECORDS COULD NOT BE COMPLETED DUE TO LACK OF LOT INFORMATION FROM THE USER FACILITY. IT IS UNKNOWN WHAT IFU INFORMATION WAS INCLUDED AT THE TIME OF THIS STUDY AND DEVICE USE. THE INSTRUCTIONS FOR USE (IFU) THAT IS CURRENTLY PACKAGED WITH THE DEVICE WAS REVIEWED FOR RELEVANT INFORMATION PERTAINING TO THE REPORTED FAILURE MODE. THE IFU DOES NOT CONTAIN INFORMATION RELATED TO THE FAILURE MODE OF OCCLUSION. THERE IS A LIMITED AVAILABILITY OF INFORMATION GIVEN THE DATE THE STUDY WAS PUBLISHED. POTENTIAL CAUSES FOR THE FAILURE MODE ARE PATIENT FACTORS, INADVERTENT USER/PROCEDURAL ISSUES, MANUFACTURING ISSUES, AND/OR DEVICE FAILURE. THE LITERATURE ARTICLE NOTES, ¿OCCLUSION APPEARED TO BE DIRECTLY RELATED TO THE VISCOSITY OF DRAINAGE, WITH PURULENT MATERIAL BEING PRESENT IN 5 OF THE 6 CASES.¿ PER THE QUALITY ENGINEERING RISK ASSESSMENT, NO FURTHER ACTION IS REQUIRED. WE WILL CONTINUE TO MONITOR FOR SIMILAR COMPLAINTS. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803 AND IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT CONTAINED HEREIN IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, NOR THAT ANY COOK DEVICE CAUSED, CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

AS REPORTED IN THE LITERATURE ARTICLE, "UROLOGIST DIRECTED PERCUTANEOUS NEPHROSTOMY TUBE PLACEMENT," A COOK COPE-LOOP NEPHROSTOMY SET BECAME OCCLUDED AFTER PLACEMENT. THE AUTHORS REPORT THE FOLLOWING: "A RETROSPECTIVE REVIEW OF 100 CONSECUTIVELY PLACED NEPHROSTOMY TUBES IN 18 PATIENTS (53 WOMEN AND 29 WOMEN) WAS DONE. ALL NEPHROSTOMY TUBES WERE PLACED BY HOUSE STAFF UNDER THE SUPERVISION OF A SINGLE ATTENDING UROLOGIST. VARIED NEPHROSTOMY TUBES WERE USED TO GAIN EXPERIENCE WITH A RANGE OF DRAINAGE SYSTEMS. "PATIENTS WERE PLACED IN THE PRONE POSITION WITH THE KNEES FLEXED AND ALL PRESSURE POINTS PADDED FOR NEPHROSTOMY TUBE INSERTION. PERCUTANEOUS ACCESS WAS ROUTINELY OBTAINED WITH A 17.5 GAUGE ANGIOMED 3-PART INITIAL PUNCTURE NEEDLE UNDER ULTRASONIC GUIDANCE WITH A 3.5 MHZ TRANSDUCER. PRELIMINARY ULTRASONIC IMAGING IDENTIFIED ADJACENT BOWEL AND OTHER INTERNAL ORGANS. KNOWLEDGE OF THE APPROXIMATE DISTANCE FROM THE SKIN TO THE COLLECTING SYSTEM HELPED ESTIMATE NEEDLE PUNCTURE DEPTH AND, IN ADDITION, MALROTATED KIDNEYS AND RENAL HILAR VESSELS WERE APPRECIATED. ULTRASONIC VISUALIZATION OF THE PUNCTURE NEEDLE WAS MAINTAINED DURING INSERTION. APPROPRIATE PLACEMENT OF THE TIP OF THE PUNCTURE NEEDLE ALSO WAS CONFIRMED BY THE EFFLUX OF URINE AND/OR INJECTION OF RADIOCONTRAST MATERIAL INTO THE COLLECTING SYSTEM. "A FLOPPY J-TIPPED GUIDE WIRE WAS PLACED INTO THE COLLECTING SYSTEM AND DOWN THE URETER WHEN POSSIBLE. FASCIAL DILATION DEVELOPED AN APPROPRIATELY SIZED TRACT. THE NEPHROSTOMY TUBE THEN WAS PLACED OVER THE GUIDE WIRE INTO THE COLLECTING SYSTEM. CONTRAST MEDIUM WAS INJECTED UNDER FLUOROSCOPIC IMAGING TO DEFINE THE ANATOMY, CONFIRM PROPER NEPHROSTOMY TUBE PLACEMENT AND EXCLUDE ANY ASSOCIATED INJURIES. ALL TUBES WERE SECURED DIRECTLY TO THE SKIN WITH 2-ZERO NYLON SUTURES AND SIMPLE 4X4 GAUZE DRESSINGS. AN ADDITIONAL FIXATION POINT WITH TAPE WAS PLACED APPROXIMATELY 20CM. FROM THE INSERTION SITE TO HELP PREVENT TWISTING AND KINKING OF THE TUBE BEFORE CONNECTION TO A CLOSED GRAVITY COLLECTION DEVICE. WHEN COPE TUBES WERE USED THE TIP WAS COILED IN THE RENAL PELVIS WITH THE PULL-STRING. BALLOON RETENTION CATHETERS WERE INFLATED WITH DILUTE CONTRAST MEDIUM TO CONFIRM POSITION. IMMEDIATE AND DELAYED COMPLICATIONS ASSOCIATED WITH NEPHROSTOMY TUBE PLACEMENT WERE IDENTIFIED, AND A DETAILED CLINICAL ASSESSMENT WAS DONE." ¿NEPHROSTOMY TUBES (100) WERE PLACED IN 29 WOMEN AND 53 MEN. A RANGE OF DIFFERENT NEPHROSTOMY TUBE ALTERNATIVES WAS USED DURING THE REVIEW PERIOD, REFLECTING A DESIRE ON OUR PART TO IDENTIFY SUPERIOR METHODS OF NEPHROSTOMY ACCESS TO LIMIT PROCEDURE RELATED AND DELAYED DRAINAGE COMPLICATIONS. WE USED 47 SINGLE J, 28 COOK-COPE, 11 RUSH NEPHROSTOMY, 9 FOLEY URINARY, 3 MALECOT AND 1 RUBBER ROBINSON URINARY CATHETERS AND 1 ARGYLE CHEST TUBE.¿ "COMPLICATIONS RELATED TO THE NEPHROSTOMY TUBES WERE ASSESSED AS BEING EITHER EARLY (INVOLVED IN CATHETER PLACEMENT AND BECOMING CLINICALLY APPARENT WITHIN 24 HOURS) OR LATE (BECOMING APPARENT MORE THAN 24 HOURS AFTER INSERTION). A TOTAL OF 27 COMPLICATIONS WAS NOTED IN 14 PATIENTS." "NEPHROSTOMY TUBE OCCLUSION OCCURRED IN 3, 10F SINGLE J CATHETERS, 1 10F COPE TUBE AND 2 FOLEY CATHETERS (16F AND 18F). OCCLUSION APPEARED TO BE DIRECTLY RELATED TO THE VISCOSITY OF DRAINAGE, WITH PURULENT MATERIAL BEING PRESENT IN 5 OF THE 6 CASES." CITED: MAHAFFEY, K.G., BOLTON, D.M., STOLLER, M.L. (1994). UROLOGIST DIRECTED PERCUTANEOUS NEPHROSTOMY TUBE PLACEMENT. THE JOURNAL OF UROLOGY, 152, 1973-1976. HTTPS://DOI.ORG/10.1016/S0022-5347(17)32282-6.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1185211 COOK COPE-LOOP NEPHROSTOMY SET LJE CATHETER, NEPHROSTOMY LJE COOK INC UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Unknown