FDA Adverse Event Malfunction Summary report: N

ASEPTIC BATTERY HOUSING

MDR report key: 13729577 · Received March 11, 2022

Report

Report Number
0008031000-2022-00004
Event Type
Malfunction
Date Received
March 11, 2022
Date of Event
February 14, 2022
Report Date
March 10, 2022
Manufacturer
ZIMMER SURGICAL SA
Product Code
MOQ
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT (B)(4). XSERIES ASEPTIC BATTERY HOUSING, PART NUMBER 89-8521-470-40, BATCH NUMBER 5014295, WAS RETURNED FOR COMPLAINT INVESTIGATION ON THE (B)(6) 2022. VISUAL AND FUNCTIONAL TESTS WERE PERFORMED. UPON RECEIPT, IT WAS CONFIRMED THAT THE LID WAS BROKEN AND THE HOLDING PIN WAS BENT. THE DEVICE WAS NOT REPAIRABLE AND HAS BEEN EXCHANGED BY A NEW ONE WITH BATCH NUMBER 5014441. DESIGN HISTORY RECORD REVIEW WAS PERFORMED AND NO ISSUE WAS DISCOVERED DURING THE MANUFACTURING PROCESS THAT COULD EXPLAIN THE DEFECT REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE ASEPTIC BATTERY HOUSING PART NUMBER 89-8521-470-40, LOT NUMBER 5014925 HAD THE BATTERY HOUSING LID SNAP OFF HALFWAY DURING SAWING. THE HOLDING PIN FOR LID IS VISIBLY BEND AND PART OF CASING IS CHIPPED. THIS EVENT IS RELATED TO A MALFUNCTION THAT COULD POTENTIALLY LEAD TO A STERILITY ISSUE. NO HARM TO PATIENT. NO ADDITIONAL SURGERY. NO DELAY. NO CONTRIBUTING CONDITION. SURGICAL TECHNIQUE FOLLOWED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1623429 ASEPTIC BATTERY HOUSING ASEPTIC BATTERY HOUSING MOQ ZIMMER SURGICAL SA NA 5014295

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other