FDA Adverse Event
Malfunction
Summary report: N
ACRA-CUT DISPOSABLE CRANIAL PERFORATOR
MDR report key: 1372634
·
Received April 15, 2009
Report
- Report Number
- 1372634
- Event Type
- Malfunction
- Date Received
- April 15, 2009
- Date of Event
- February 27, 2009
- Report Date
- April 15, 2009
- Manufacturer
- ACRA-CUT, INC.
- Product Code
- HBF
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MI, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
WHILE DRILLING A BURR HOLE WITH THE DRILL, THE 14MM BURR BROKE OFF AT THE SHAFT ENTRY TO THE DRILL HANDLE. A NEW DRILL AND NEW 14 MM BURR WERE OBTAINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACRA-CUT DISPOSABLE CRANIAL PERFORATOR | BLADE, CRANIAL | HBF | ACRA-CUT, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |