FDA Adverse Event Injury Summary report: N

INX FOR INLINE WITH DWL

MDR report key: 13724974 · Received March 10, 2022

Report

Report Number
1523574-2022-00002
Event Type
Injury
Date Received
March 10, 2022
Date of Event
February 26, 2022
Report Date
July 11, 2022
Manufacturer
FERNO-WASHINGTON, INC.
Product Code
FPO
UDI-DI
00190790001339
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE STRETCHER WAS RETURNED TO MANUFACTURER AND A VISUAL AND FUNCTIONAL EVALUATION WAS CONDUCTED. THERE WERE NO OBSERVATIONS OF MECHANICAL ISSUES THAT WOULD CONTRIBUTE TO THE STRETCHER NOT ENGAGING WITH THE HOOK. THERE WERE NO OBSERVATIONS OF STRIKE MARKS FROM A DROP AND THE SAFETY BAIL SWUNG FREELY THROUGH ITS MOTION AND ENGAGED WITH HOOK EVERY TIME DURING THE UNLOADING CYCLE. IT IS SUSPECTED THE EVENT WAS ATTRIBUTED TO UNINTENTIONAL USE ERROR.

Description of Event or Problem · 0

THE CUSTOMER REPORTED WHILE UNLOADING A PATIENT FROM THE AMBULANCE, THE SAFETY BAIL ON THE HEAD END OF THE STRETCHER ALLEGEDLY DID NOT ENGAGE WITH THE SAFETY HOOK THERFORE; THE STRETCHER FELL TO THE BUMPER AND TIPPED OVER. THE PATIENT WAS REPORTED TO HAVE POSSIBLY HIT THEIR HEAD AND WAS TREATED IN THE ER FOR AN ABRASION. THE STRETCHER WILL BE EVALUATED AT FERNO. NO FURTHER DETAILS WERE PROVIDED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED WHILE UNLOADING A PATIENT FROM THE AMBULANCE, THE SAFETY BAIL ON THE HEADEND OF THE STRETCHER ALLEGEDLY DID NOT ENGAGE WITH THE SAFETY HOOK AND THE STRETCHER FELL TO THE BUMPER AND TIPPED OVER. THE PATIENT WAS REPORTED TO HAVE POSSIBLY HIT THEIR HEAD AND WAS TREATED IN THE ER FOR AN ABRASION. THE STRETCHER WILL BE EVALUATED AT FERNO. NO FURTHER DETAILS WERE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2117599 INX FOR INLINE WITH DWL INX FOR INLINE WITH DWL FPO FERNO-WASHINGTON, INC. 0015811 00190790001339

Patients

Seq Age Sex Outcome Treatment
1 89 YR Male Other