FDA Adverse Event Malfunction Summary report: N

PUMP MMT-1714K 630G BLACK MMOL CANADA

MDR report key: 13723919 · Received March 10, 2022

Report

Report Number
2032227-2022-125798
Event Type
Malfunction
Date Received
March 10, 2022
Date of Event
December 1, 2021
Report Date
June 14, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
UDI-DI
000000763000315702
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. (B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Additional Manufacturer Narrative · 0

RETAINER RING=BLACK. DEVICE WAS RETURNED FOR PUMP ERROR 53 AND PUMP ERROR 4 ALARMS FOUND ON DEC 01, 2021. ALLEGATION TO HIGH BGS NOTED. DEVICE'S HISTORY AND TRACE FILES DOWNLOADED SUCCESSFULLY USING THUS SOFTWARE. UNABLE TO PERFORM DISPLACEMENT TEST AND SELF TEST DUE TO UNEXPECTED INSULIN PUMP RESETS CAUSING PUMP ERROR 53 ALARMS. THE HISTORY/TRACE DOWNLOAD CONFIRMED PUMP ERROR 4 ON 12/01/2021 AT 10:02AM. PUMP ERROR 53(LINE NUMBER 24582 FILE NUMBER 32216) ALARMS CONFIRMED ON THE PUMP HISTORY/TRACE FILES ON 12/01/2021 AT 9:35AM. ¿PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND¿NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE¿ON THE ELECTRONIC ASSEMBLY, MOTOR, OR FORCE SENSOR. TEST P-CAP / RESERVOIR LOCKS PROPERLY INTO PLACE. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: SCRATCHED CASE. PUMP ERROR 4 ALARMS CONFIRMED IN THE ON 12/01/2021 AT 10:02AM AND PUMP ERROR 53(LINE NUMBER 24582 FILE NUMBER 32216) ALARMS CONFIRMED ON THE PUMP HISTORY/TRACE FILES ON 12/01/2021 AT 9:35AM DUE TO A POSSIBLE HARDWARE ERROR (ESF#1925624). PROBLEM ISOLATED TO THE ELECTRONIC ASSEMBLIES. UNABLE TO VERIFY CUSTOMER ALLEGED FOR HIGH BGS. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION THAT PUMP ERROR 53, PUMP ERROR 4 OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2360229 PUMP MMT-1714K 630G BLACK MMOL CANADA ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1714K HG5TM05 000000763000315702

Patients

Seq Age Sex Outcome Treatment
1 Unknown