FDA Adverse Event
Malfunction
Summary report: N
OMNIPOD INSULIN PUMP
MDR report key: 1372291
·
Received February 19, 2009
Report
- Report Number
- 3004464228-2009-00115
- Event Type
- Malfunction
- Date Received
- February 19, 2009
- Date of Event
- January 25, 2009
- Report Date
- January 25, 2009
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE RETURNED PRODUCT EVAL CONFIRMED AN INTERNAL LEAK WHICH CAUSED DAMAGE TO THE DEVICE AFTER BEING FILLED WITH INSULIN. THE USER IS INSTRUCTED IN THE USER GUIDE TO FREQUENTLY MONITOR THEIR BG LEVELS. BY FOLLOWING THIS RECOMMENDATION, THE USER WAS AWARE OF HIGH LEVELS AND WAS ABLE TO START A NEW POD SUCCESSFULLY. THE LOT WAS FOUND TO HAVE MET ALL ACCEPTANCE CRITERIA.
Description of Event or Problem · 1
CUSTOMER'S WIFE CALLED TO REPORT THAT HER CUSTOMER'S POD WAS WORN FOR APPROXIMATELY 12 HOURS AND REMOVED DUE TO ELEVATED BLOOD GLUCOSE LEVELS (BGS). HIS BGS RANGED BETWEEN 162-341 MG/DL BEFORE TAKING THE POD OFF AND STARTING A NEW ONE. NO FURTHER ISSUES WERE IDENTIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L30074 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Other |