FDA Adverse Event
Injury
Summary report: N
LINOX TD 65/16
MDR report key: 1369761
·
Received April 24, 2009
Report
- Report Number
- 1028232-2009-00436
- Event Type
- Injury
- Date Received
- April 24, 2009
- Date of Event
- March 12, 2009
- Report Date
- March 30, 2009
- Manufacturer
- BIOTRONIK GMBH AND CO.
- Product Code
- DTB
- PMA / PMN Number
- P980023
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THEREFORE, THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE WERE RE-INVESTIGATED. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING PRODUCTION AND FINAL ACCEPTANCE TEST. THE BIOTRONIK STERILIZATION PROTOCOL CONFIRMED THAT ALL STERILIZATION PARAMETERS, SUCH AS GAS CONCENTRATION, TEMPERATURE, HUMIDITY, ETC., WERE IN THEIR SPECIFIED RANGES. ALSO, THE INVESTIGATION OF THE MICROBIOLOGICAL INDICATORS SHOWED THE SUCCESSFUL COMPLETION OF THE STERILIZATION PROCESS. IN SUMMARY, THE INFECTION WAS NOT DEVICE RELATED.
Description of Event or Problem · 1
THIS SYSTEM WAS REMOVED DUE TO INFECTION AND EROSION. LUMAX 340 DR-T, MDR: 1028232-2009-00434. SETROX S 53, MDR: 1028232-2009-00435.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LINOX TD 65/16 | PACER LEAD | DTB | BIOTRONIK GMBH AND CO. | 351337 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Hospitalization |