FDA Adverse Event Injury Summary report: N

IMPL TWIST MP-1 5.0 MM 10 MM

MDR report key: 13695270 · Received March 8, 2022

Report

Report Number
0002023141-2022-00605
Event Type
Injury
Date Received
March 8, 2022
Date of Event
July 26, 2021
Report Date
March 8, 2022
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K013494
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT (B)(4). ADDITIONAL 510(K) NUMBER IS K962106. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE # 19 LOST INTEGRATION DUE TO PERI-IMPLANTITIS AT THE IMPLANT SITE AND WAS REMOVED. THE SITE WAS GRAFTED AND A NEW IMPLANT WAS PLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
936750 IMPL TWIST MP-1 5.0 MM 10 MM DENTAL IMPLANT DZE ZIMMER DENTAL 991673

Patients

Seq Age Sex Outcome Treatment
1 70 YR Female Required Intervention