FDA Adverse Event
No answer provided
Summary report: N
DRAGER RADIANT WARMER BED TEMPERATURE PROBE
MDR report key: 13691863
·
Received March 8, 2022
Report
- Report Number
- 13691863
- Event Type
- No answer provided
- Date Received
- March 8, 2022
- Date of Event
- December 24, 2021
- Report Date
- February 22, 2022
- Manufacturer
- DRAEGERWERK AG & CO. KGAA
- Product Code
- FLL
- Report Source
- User Facility report
- Reporter Location
- PA
- Reporter Occupation
- RISK MANAGER
- Health Professional
- N
Narratives
Description of Event or Problem · 0
NOTICED INFANT WAS WARM IN AFTERNOON CARE. RN TRIED TO WEAN THE BED TEMPERATURE DOWN. INFANT WAS WARMER AT NEXT CARE TIME. RN ATTEMPTED TO WEAN BED TEMPERATURE A SECOND TIME. RN CHECKS TEMP AN HOUR LATER, TEMPERATURE IS 101.4 AND TEMP PROBE IS ON CORRECTLY BUT READING COLD AND BED IS AT HIGHEST HEAT INTENSITY. PRODUCT SAVED FOR PRODUCTS NURSE. MANUFACTURER AWARE OF ISSUE. NEW PROBE USED, ISSUE DID NOT REPEAT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1496151 | DRAGER RADIANT WARMER BED TEMPERATURE PROBE | THERMOMETER, ELECTRONIC, CLINICAL | FLL | DRAEGERWERK AG & CO. KGAA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |