FDA Adverse Event
Injury
Summary report: N
GENTLEPOWER LUX CONTRA ANGLE 25LPA
MDR report key: 1368883
·
Received April 27, 2009
Report
- Report Number
- 8010493-2009-00007
- Event Type
- Injury
- Date Received
- April 27, 2009
- Date of Event
- June 5, 2008
- Report Date
- June 5, 2008
- Manufacturer
- KALTENBACH & VOIGT GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
PENICILLIN WAS PRESCRIBED TO THE PATIENT FOR TREATMENT OF THE BURN. THE HANDPIECE WAS SENT IN FOR EVALUATION. AT THE EVALUATION, IT WAS FOUND THAT THE TOP OF THE HANDPIECE WAS DENTED AT THE BACK CAP. THE DENT WAS CAUSING THE BACK CAP TO BE DEPRESSED WHICH CAUSED THE HANDPIECE TO OVERHEAT. A REPLACEMENT HANDPIECE WAS SENT TO THE DOCTOR.
Description of Event or Problem · 1
THE DOCTOR WAS WORKING ON BRIDGE TOOTH #7 THEN NOTICED A BURN ON THE PATIENT¿S CHEEK AND TONGUE. BURN DIAMETER WAS DIME SIZE ON THE CHEEK AND LARGER THEN QUARTER SIZE ON THE TONGUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX CONTRA ANGLE 25LPA | DENTAL HANDPIECE AND ACCESSORIES | EFB | KALTENBACH & VOIGT GMBH | 25LPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |