FDA Adverse Event Injury Summary report: N

LINOX TD 65-16

MDR report key: 1367963 · Received April 20, 2009

Report

Report Number
1028232-2009-00431
Event Type
Injury
Date Received
April 20, 2009
Date of Event
March 11, 2009
Report Date
March 19, 2009
Manufacturer
BIOTRONIK GMBH AND CO.
Product Code
LWS
PMA / PMN Number
P980023
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THEREFORE, THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE WERE RE-INVESTIGATED. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING PRODUCTION AND FINAL ACCEPTANCE TEST. THE BIOTRONIK STERILIZATION PROTOCOL CONFIRMED THAT ALL STERILIZATION PARAMETERS, SUCH AS GAS CONCENTRATION, TEMPERATURE, HUMIDITY, ETC., WERE IN THEIR SPECIFIED RANGES. ALSO, THE INVESTIGATION OF THE MICROBIOLOGICAL INDICATORS SHOWED THE SUCCESSFUL COMPLETION OF THE STERILIZATION PROCESS. IN SUMMARY, THE INFECTION WAS NOT DEVICE RELATED.

Description of Event or Problem · 1

PER OOS, THIS SYSTEM WAS REMOVED DUE TO INFECTION. LUMAX 340 DR-T, MDR 1028232-2009-00429. SETROX S 45, MDR 1028232-2009-00430.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LINOX TD 65-16 ICD LEAD LWS BIOTRONIK GMBH AND CO. 351337

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization