FDA Adverse Event
Injury
Summary report: N
T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
MDR report key: 13673149
·
Received March 4, 2022
Report
- Report Number
- 3013756811-2022-21464
- Event Type
- Injury
- Date Received
- March 4, 2022
- Date of Event
- February 12, 2022
- Report Date
- February 12, 2022
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613762
- PMA / PMN Number
- K201214
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
Additional Manufacturer Narrative · 0
THE DEVICE HAS BEEN RECEIVED FOR EVALUATION; HOWEVER, DEVICE EVALUATION IS NOT YET COMPLETE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION.
Description of Event or Problem · 0
IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. ADDITIONALLY, IT WAS REPORTED THAT A CARTRIDGE ALARM OCCURRED DURING BASAL DELIVERY. IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ELEVATED BLOOD GLUCOSE (BG) LEVEL OF 200-533 MG/DL. CAUSE WAS DUE TO MALFUNCTION INTERRUPTING INSULIN DELIVERY. A MANUAL INJECTION WAS ADMINISTERED TO ADDRESS BG. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1532171 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | 00850006613762 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR | Male | Other |