FDA Adverse Event Malfunction Summary report: N

WALLACH ULTRA FREEZE

MDR report key: 13663850 · Received March 3, 2022

Report

Report Number
1216677-2022-00058
Event Type
Malfunction
Date Received
March 3, 2022
Date of Event
January 20, 2022
Report Date
March 3, 2022
Manufacturer
COOPERSURGICAL, INC.
Product Code
GEH
PMA / PMN Number
K935010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

COOPERSURGICAL, INC. IS CURRENTLY INVESTIGATING THE REPORTED CONDITION.

Additional Manufacturer Narrative · 0

INVESTIGATION. X-REVIEW DHR X-INSPECT RETURNED SAMPLES. *ANALYSIS AND FINDINGS COMPLAINT (B)(4). DISTRIBUTION HISTORY: THIS COMPLAINT UNIT WAS MANUFACTURED AT CSI ON 09/01/2015 UNDER WO #172173 AND SHIPPED ON 10/07/2015. MANUFACTURING RECORD REVIEW: DHR 172173 WAS REVIEWED AND NO NON-CONFORMITIES, RELATED TO THE COMPLAINT CONDITION, WERE NOTED. INCOMING INSPECTION REVIEW: NOT APPLICABLE. SERVICE HISTORY RECORD: THIS UNIT WAS SERVICED IN APRIL 2016 FOR A BOTTLE THAT HAD LOST ITS VACUUM SEAL. HISTORICAL COMPLAINT REVIEW: A REVIEW OF THE 2-YEAR COMPLAINT HISTORY SHOWED SIMILAR REPORTED COMPLAINT CONDITION. PRODUCT RECEIPT: THE COMPLAINT UNIT WAS RETURNED ON REPAIR LOG 97786. VISUAL EVALUATION: VISUAL EXAMINATION OF THE COMPLAINT UNIT REVEALED PHYSICAL DAMAGE TO THE TRIGGER. FUNCTIONAL EVALUATION: COMPLAINT UNIT WAS FUNCTIONALLY EVALUATED AND FOUND NOT TO FUNCTION PROPERLY. ROOT CAUSE: THE BROKEN TRIGGER IS BEING ATTRIBUTED TO HANDLING ERROR BY THE END USER. *CORRECTION AND/OR CORRECTIVE ACTION THE UNIT WAS REPAIRED, TESTED AND RETURNED TO THE CUSTOMER. NO FURTHER CORRECTIVE ACTION IS NECESSARY. *PREVENTATIVE ACTION ACTIVITY COOPERSURGICAL WILL CONTINUE TO MONITOR THIS COMPLAINT CONDITION FOR TRENDS.

Description of Event or Problem · 0

BROKEN HANDLE. CONFIRMED COMPLAINT: TRIGGER BROKEN OFF. REPAIR ORDER 97786. WALLACH ULTRA FREEZE 900076 E-COMPLAINT(B)(4).

Description of Event or Problem · 0

BROKEN HANDLE. CONFIRMED COMPLAINT: TRIGGER BROKEN OFF. REPAIR ORDER 97786. 1216677-2022-00058 WALLACH ULTRA FREEZE 900076 E-COMPLAINT-(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
899949 WALLACH ULTRA FREEZE WALLACH ULTRA FREEZE GEH COOPERSURGICAL, INC. 900076 N/A

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other