FDA Adverse Event
Malfunction
Summary report: N
OMNIPOD INSULIN PUMP
MDR report key: 1366042
·
Received January 23, 2009
Report
- Report Number
- 3004464228-2009-00052
- Event Type
- Malfunction
- Date Received
- January 23, 2009
- Date of Event
- January 2, 2009
- Report Date
- January 2, 2009
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE RETURNED PRODUCT EVAL CONFIRMED AN INTERNAL LEAK WHICH CAUSED DAMAGE TO THE DEVICE AFTER BEING FILLED WITH INSULIN. THE USER IS INSTRUCTED IN THE USER GUIDE TO FREQUENTLY MONITOR THEIR BG LEVELS. BY FOLLOWING THIS RECOMMENDATION, THE USER WAS AWARE OF HIGH BG LEVELS AND WAS ABLE TO START A NEW POD SUCCESSFULLY. THE LOT WAS FOUND TO HAVE MET ALL ACCEPTANCE CRITERIA.
Description of Event or Problem · 1
CUSTOMER'S MOTHER CALLED TO REPORT A POD ALARMED WHILE HER SON WAS SLEEPING. SHE ALSO STATED THAT HER SON'S BLOOD GLUCOSE READINGS WERE ELEVATED AND RANGED BETWEEN 230-445 MG/DL. CUSTOMER WAS ABLE TO START A NEW POD SUCCESSFULLY. NO FURTHER ISSUES WERE IDENTIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L11850 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR | Other |