FDA Adverse Event Malfunction Summary report: N

CERAMO FEHLING PUNCH

MDR report key: 13657803 · Received March 3, 2022

Report

Report Number
9615005-2022-80006
Event Type
Malfunction
Date Received
March 3, 2022
Date of Event
January 17, 2022
Report Date
March 2, 2022
Manufacturer
FEHLING INSTRUMENTS GMBH & CO. KG
Product Code
HAE
UDI-DI
04058749063394
PMA / PMN Number
K153243
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

BASED ON THE LOT NUMBER, WE HAVE CHECKED THE DHR: NO FINDINGS WE WILL EXAMINE THE DEVICE AS SOON AS POSSIBLE.

Additional Manufacturer Narrative · 0

BASED ON THE LOT NUMBER, WE HAVE CHECKED THE DHR: NO FINDINGS. THE FOOT OF THE PUNCH BROKE OFF, BUT THE BROKEN PIECE WASN'T RETURNED. MICROSCOPE PHOTOS SHOW SERIOUS DEFORMATIONS OF THE CUTTING EDGE OF THE SLIDER. BASED ON THESE FINDINGS, WE ASSUME THAT THE BREAKAGE WAS CAUSED BY OVERLOADING DUE TO CUTTING TOO HARD MATERIAL. (SEE ENCLOSED PICTURES) THE PUNCH IS MADE OF HARDENED STEEL AND IS USED FOR PUNCHING BONE STRUCTURES THINNED OUT BY MILLING. WHEN USED PROPERLY, IT IS NOT POSSIBLE FOR LIPPAGE OF THIS TYPE TO OCCUR ON THE CUTTING EDGE. A MATERIAL OR MANUFACTURING DEFECT CAN BE RULED OUT DUE TO THE HOMOGENEOUS STRUCTURE OF THE FRACTURE SURFACE.

Description of Event or Problem · 0

A CERVICAL KERRISION BROKE DURING THE PROCEDURE WHILE IN USE. ALL PIECES WER RETRIEVED BY THE SURGEON.

Description of Event or Problem · 0

A CERVICAL KERRISION BROKE DURING THE PROCEDURE WHILE IN USE. ALL PIECES WER RETRIEVED BY THE SURGEON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
653937 CERAMO FEHLING PUNCH EJECTOR PUNCH HAE FEHLING INSTRUMENTS GMBH & CO. KG NNR-0F X17J543WM 04058749063394

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention