FDA Adverse Event Malfunction Summary report: N

NEUROMONICS OASIS TINNITUS TREATMENT DEVICE

MDR report key: 1365389 · Received February 3, 2009

Report

Report Number
3005777056-2009-00001
Event Type
Malfunction
Date Received
February 3, 2009
Date of Event
January 6, 2009
Report Date
February 3, 2009
Manufacturer
NEUROMONICS INC.
Product Code
KLW
PMA / PMN Number
K043274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP WITH THE PT INDICATED NO MEDICAL INTERVENTION, SERIOUS INJURY, OR SUBSEQUENT EFFECTS FROM THE INCIDENT. IGNITION IS BELIEVED TO HAVE BEEN CAUSED BY A THERMAL RUNAWAY OF THE BATTERY WHEN CHARGING WAS INITIATED. THE PT HAD NOTED THAT THE UNIT HAD BEEN CHARGED ALMOST DAILY DURING THE PREVIOUS 18 MONTHS OF USAGE WITHOUT ANY ISSUES. INVESTIGATION TO DATE HAS INCLUDED: REVIEW AND ANALYSIS OF THE MALFUNCTIONING UNIT IN CONJUNCTION WITH THE BATTERY MANUFACTURER; REVIEW OF DESIGN; REVIEW OF VERIFICATION TESTS, ASSEMBLY PROCEDURES, AND MANUFACTURING RECORDS FOR THE MALFUNCTIONING UNIT AND OTHER UNITS. THIS CURRENTLY APPEARS TO BE AN ISOLATED EVENT CONSIDERING THE LARGE NUMBER OF UNITS MANUFACTURED WITHOUT SIMILAR INCIDENTS, AND CIRCUITRY/PHYSICAL DESIGN INTENDED TO MITIGATE RISKS ASSOCIATED WITH THE USE OF RECHARGEABLE BATTERIES. THE REVIEW PROCESS CONTINUES IN CONJUNCTION WITH THE BATTERY SUPPLIER, CONTRACT MANUFACTURER AND PRODUCT DESIGNER EFFORTS TO IDENTIFY POTENTIAL ROOT CAUSES AND EVALUATE THE RISK OF RECURRENCE. SUPPLEMENTARY INFO WILL BE PROVIDED IF FURTHER PERTINENT INFO IS IDENTIFIED.

Description of Event or Problem · 1

PT PLUGGED IN THE DEVICE TO CHARGE IT AND PLACED IT ON A GRANITE BENCHTOP. THE DEVICE WAS CONTAINED IN ITS CARRYING POUCH. THE UNIT OVERHEATED RAPIDLY AND BEGAN TO EMIT FLAMES. PT THREW THE DEVICE IN A NEARBY SINK TO EXTINGUISH THE FLAMES AND REPORTED THAT HIS THUMB WAS BLISTERED IN THE PROCESS. PT DID NOT ELECT TO SEEK MEDICAL TREATMENT FOR THE BLISTERING. THE DEVICE WAS SIGNIFICANTLY DAMAGED AND THERE WAS SOME SMOKE PRODUCED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEUROMONICS OASIS TINNITUS TREATMENT DEVICE OASIS TINNITUS TREATMENT DEVICE KLW NEUROMONICS INC. 110 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK