I-STAT CG8+ CARTRIDGE
Report
- Report Number
- 2245578-2022-00032
- Event Type
- Malfunction
- Date Received
- March 1, 2022
- Date of Event
- February 15, 2022
- Report Date
- April 5, 2022
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- CHL
- UDI-DI
- 10054749000163
- PMA / PMN Number
- K940918
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 01-APR-2022. THROUGH INVESTIGATION OF A PREVIOUS CUSTOMER COMPLAINT (INCIDENT: 877194), A DEFICIENCY FOR ELEVATED RATE OF UNEXPECTED (HIGH) ICA RESULTS WHEN TESTING WITH CARTRIDGES FROM CG8+ LOT W21198 WAS IDENTIFIED. THIS ISSUE IS UNDER INVESTIGATION UNDER QUALITY RECORDS (QRS) 775835 AND 796335.
ON 24-FEB-2022, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CG8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT IONIZED CALCIUM RESULTS OF 1.59 & 2022 ON A PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD :I-STAT , DATE :(B)(6)2022 10:19 , TESTED RESULT : 1.59 MMOL/L , AND SAMPLE : A WB. METHOD :I-STAT ,DATE :(B)(6)2022 11:53 , TESTED RESULT : 2.22 MMOL/L , AND SAMPLE : B WB. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
NA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1793376 | I-STAT CG8+ CARTRIDGE | CG8+ CARTRIDGE | CHL | ABBOTT POINT OF CARE | NA | W21198 | 10054749000163 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |