FDA Adverse Event Injury Summary report: N

CHLORAPREP ONE STEP 1.5 ML FREPP

MDR report key: 13625943 · Received February 28, 2022

Report

Report Number
3004932373-2022-00058
Event Type
Injury
Date Received
February 28, 2022
Date of Event
December 15, 2021
Report Date
May 4, 2022
Manufacturer
CAREFUSION 213, LLC 0113
Product Code
OJU
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

(B)(6) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

(B)(4). INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4).

Additional Manufacturer Narrative · 0

BATCH RECORD FOR PN 930299 LN 1246784 WAS REVIEWED AND THERE WERE NO NON-CONFORMANCES RELATED TO THE DEFECT OF "GLASS ON FOAM OR PLASTIC BODY" OR "OPEN SEAL" DURING THE MANUFACTURING OF THE LOT. WITHOUT THE ACTUAL SAMPLE FOR ANALYSIS THE ROOT CAUSE COULD NOT BE CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE NARRATIVE

Description of Event or Problem · 0

MATERIAL# 930299. BATCH# 1246784. IT WAS REPORTED SHARPS: VERBATIM: :DESCRIBE THE EVENT OR PROBLEM: ADULT FEMALE BEING PREPARED FOR AN IV PLACEMENT BEFORE HER BILATERAL BREAST MRI EXAM. AS THE TECHNOLOGIST WAS CLEANING THE SKIN THE PATIENT WAS SCRAPED BY SOMETHING POSSIBLY INSIDE/ON THE CHLORHEXIDINE SKIN PREP - LEFT ANTECUBITAL/FOREARM AREA - THERE WAS SKIN EROSION SUBSTANTIAL ENOUGH TO CAUSE A BLEED. PATIENT WAS ASSESSED BY RADIOLOGY RN AND SKIN WAS CLEANED & ANTISEPTIC AROUND THE AFFECTED AREA AND DRESSING APPLIED. NO DELAY TO PROCEDURE. WHAT WAS THE ORIGINAL INTENDED PROCEDURE? : PATIENT PREOPERATIVE SKIN PREPARATION: SITE-PREP FOR IV PLACEMENT.

Description of Event or Problem · 0

MATERIAL# 930299. BATCH# 1246784. IT WAS REPORTED SHARPS: VERBATIM: DESCRIBE THE EVENT OR PROBLEM: ADULT FEMALE BEING PREPARED FOR AN IV PLACEMENT BEFORE HER BILATERAL BREAST MRI EXAM. AS THE TECHNOLOGIST WAS CLEANING THE SKIN THE PATIENT WAS SCRAPED BY SOMETHING POSSIBLY INSIDE/ON THE CHLORHEXIDINE SKIN PREP - LEFT ANTECUBITAL/FOREARM AREA - THERE WAS SKIN EROSION SUBSTANTIAL ENOUGH TO CAUSE A BLEED. PATIENT WAS ASSESSED BY RADIOLOGY RN AND SKIN WAS CLEANED & ANTISEPTIC AROUND THE AFFECTED AREA AND DRESSING APPLIED. NO DELAY TO PROCEDURE. WHAT WAS THE ORIGINAL INTENDED PROCEDURE? : PATIENT PREOPERATIVE SKIN PREPARATION: SITE-PREP FOR IV PLACEMENT.

Description of Event or Problem · 0

MATERIAL# 930299 BATCH# 1246784 IT WAS REPORTED SHARPS VERBATIM DESCRIBE THE EVENT OR PROBLEM: ADULT FEMALE BEING PREPARED FOR AN IV PLACEMENT BEFORE HER BILATERAL BREAST MRI EXAM. AS THE TECHNOLOGIST WAS CLEANING THE SKIN THE PATIENT WAS SCRAPED BY SOMETHING POSSIBLY INSIDE/ON THE CHLORHEXIDINE SKIN PREP - LEFT ANTECUBITAL/FOREARM AREA - THERE WAS SKIN EROSION SUBSTANTIAL ENOUGH TO CAUSE A BLEED. PATIENT WAS ASSESSED BY RADIOLOGY RN AND SKIN WAS CLEANED & ANTISEPTIC AROUND THE AFFECTED AREA AND DRESSING APPLIED. NO DELAY TO PROCEDURE. WHAT WAS THE ORIGINAL INTENDED PROCEDURE? : PATIENT PREOPERATIVE SKIN PREPARATION: SITE-PREP FOR IV PLACEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
943917 CHLORAPREP ONE STEP 1.5 ML FREPP 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL OJU CAREFUSION 213, LLC 0113 1246784

Patients

Seq Age Sex Outcome Treatment
1 32 YR Female Other