FDA Adverse Event Malfunction Summary report: N

S/5 CRITICAL CARE MONITOR

MDR report key: 1362534 · Received March 24, 2009

Report

Report Number
9610105-2009-00013
Event Type
Malfunction
Date Received
March 24, 2009
Date of Event
July 14, 2008
Report Date
March 24, 2009
Manufacturer
GE HEALTHCARE FINLAND OY
Product Code
MHX
PMA / PMN Number
K071889
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B) (6) PRIVACY LAW PROHIBITS RELEASING PATIENT INFORMATION. GE HEALTHCARE INVESTIGATED THE NMF CPU MOTHERBOARD AND F-CPU POWER BOARD. BOTH BOARDS WERE BADLY BURNED. AS A RESULT, THE BOARDS WERE UNABLE TO BE REPAIRED OR TESTED. VISUAL INSPECTION REVEALED THAT A CAPACITOR AND PRINTED CIRCUIT BOARD AT THE SAME LOCATION BENEATH THE CAPACITOR WAS BURNED, RESULTING IN A HOLE IN THE PRINTED CIRCUIT BOARD ON THE NMF CPU MOTHERBOARD. SEVERAL COMPONENTS WERE BURNED ON THE F-CPU POWER BOARD, RESULTING IN A HOLE IN THE PRINTED CIRCUIT BOARD AS WELL. DUE TO THE EXTENT OF THE DAMAGE TO BOTH BOARDS, THE ROOT CAUSE FOR THE COMPONENT FAILURES REMAINS UNKNOWN, AS NEITHER BOARD COULD BE TESTED. THE DEFECTIVE BOARDS WERE REPLACED AND THE MONITOR WAS PLACED BACK INTO USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT PARTS OF THE POWER BOARD AND CPU MOTHERBOARD FAILED ON THE S/5 CRITICAL CARE MONITOR, RESULTING IN A LOSS OF MONITORING WITHOUT AN ALARM. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 S/5 CRITICAL CARE MONITOR MODULAR MONITORS MHX GE HEALTHCARE FINLAND OY S/5 CCM